Operating Instructions DWM 2000 D 

 

 
 

Fct. n° 

Text 

Description and settings

2.2.0. 

CURRENT OUT 

Sub menu 2.2.0. Current output

2.2.1. CURRENT? 

Not 

available 

2.2.2. 

I 0% 

Calibration of the current output for i

0%

 

Measure the exact value on a milliammeter and press the “+” or 
“-” key in order to obtain the wished value for i

0%

 

3 mA 

 i

0%

 

 12 mA 

2.2.3. 

I 100% 

Calibration of the current output for i

100%

 

Measure the exact value on a milliammeter and press the “+” or 
“-” key in order to obtain the wished value for i

100%

 

 12 mA 

 i

100%

 

 21 mA 

2.2.4. 

TIME CONST. 

Time constant for output of the measured values 
Range: 5, 10, 15, 20, 25, 30, 50 m 

2.4.0. 

SPECIAL 

Sub menu 2.4.0. Special functions

2.4.1. 

LANGUAGE 

Language for display text 
“GB”: English 
“F”: French 
“D”: German 

2.4.2. PASSWORD 

Not 

available 

2.4.3. 

EEP PARAM. 

Display of the different parameters memorised in the 
EEPROM: see section “2.4 Parameters Stored in the 
EEPROM”. Read only. 

2.4.4. 

FILTER 

Activation of an electronic filter for noisy applications 
(foam, solid contents). 
Select “YES” or “NO”, default setting “YES”. 

2.4.5. DISPLAY 

Display 

of 

the actual velocity in m/s. The velocity is displayed 

after quitting the programming menu. 
Select “YES” or “NO”. Default setting “NO”. 
It must be programmed to “NO” before you disconnect the 
DWM 2000 D. 

2.4.6. 

DIAMETER 

Diameter of the pipe into which the sensor is inserted. This 
value is needed for calculating the flow rate. If you enter a 
value of “0”, this switches off flow rate on the indicator display. 
Refer also to the CAUTION that follows. 
50 mm 

 diameter 

 XXXX mm 

 
 

 

CAUTION 

Do not use menu 2.4.6 to read the setting for the pipe diameter. This will reset 
the value to default “0” and switch off the flow rate line on the indicator 
display. Use menu 2.4.3 to read EEPROM settings (this includes DIAMETER). 

If the value is reset to “0”, you must re-enter the pipe diameter in menu 2.4.6. 

 
 
 
 

Summary of Contents for DWM 2000 D

Page 1: ...DWM 2000 D DWM 2000 Electromagnetic Flowmeter with LCD Indicator Operating instructions...

Page 2: ...mable parameters 4 2 1 1 Flow calibration 4 2 1 2 Current output adjustment 4 2 1 3 Time constant 4 2 2 Electronics module checks 5 2 3 Programming structure Software n 1 02 5 2 3 1 User interface but...

Page 3: ...Operating Instructions DWM 2000 D 3 1 Display data in operating mode...

Page 4: ...a Vm with Va actual velocity Vm measured velocity reference value 2 1 2 Current output adjustment The minimum value i0 and the maximum value i100 of the current output at normal operating conditions c...

Page 5: ...and the corrective actions to be undertaken Call up the error messages menu 1 2 2 and note the last one Refer to the error message list Replace the electronics module in case of fatal error deactivate...

Page 6: ...6 Operating Instructions DWM 2000 D 2 3 2 Menu navigation Go through the steps given in the illustrations that follow to get to the required menu...

Page 7: ...Operating Instructions DWM 2000 D 7...

Page 8: ...eset the instrument to zero Refer to the installation manual to use the Zero Cal button on the DWM 2000 electronics block correctly 1 2 0 DIAGNOSTIC Sub menu 1 2 0 Diagnostic 1 2 1 ALARM MODE Deactiva...

Page 9: ...orised in the EEPROM see section 2 4 Parameters Stored in the EEPROM Read only 2 4 4 FILTER Activation of an electronic filter for noisy applications foam solid contents Select YES or NO default setti...

Page 10: ...test level CUSTOM CORRECTION YES NO Indication of activation of low velocity linearization YES EP CHECKS Result of the data check in EPROM see menu 1 1 6 Value 22309 With soft V1 02 FS Display of the...

Page 11: ...or 1 m s Replace the electronics module ZERO KEY Dysfunction of ZERO key short circuit Replace the electronics module EP CHECK Data loss in EPROM software Replace the electronics module EEP CHECK Data...

Page 12: ...e 7 0 095 911 7165 Fax 7 0 095 742 8873 krohne dol ru India Krohne Marshall Ltd A 34 35 M I D C Industrial Area H Block Pimpri Poona 411018 Phone 91 0 202 744 2020 Fax 91 0 202 744 2020 pcu vsnl net I...

Reviews: