KROHNE DWM 2000 D Operating Instructions Manual Download Page 10

 
 

10  

 

Operating Instructions DWM 2000 D 

 

 

 

2.4 

Parameters stored in the EEPROM (menu 2.4.3) 

 

Parameter Comment 

Typical 

value

CHECKS 1 

Check EEPROM n°1 

               - 

CHECKS 2 

Check EEPROM n°2 

               - 

CHECKS BYTE 1 

Check EEPROM 

               - 

CHECKS BYTE 2 

Check EEPROM 

               - 

CPT ALARM. 

Counting of all the error messages 
since the first power-up. 

               - 

CPT ALARM.2: 

Counting of all the error messages 
since the last reset.  

               - 

TEST 

Indication of auto diagnostic test 
level 

CUSTOM 

CORRECTION 
YES/NO 

Indication of activation of low velocity 
linearization. 

YES 

EP CHECKS 

Result of the data check in EPROM 
(see menu 1.1.6.) 

Value = 22309 
With soft V1.02 

FS 

Display of the programmed full scale 
values (see menu 1.1.2.) 

1 m/s ... 8 m/s 

GK 

Primary head calibration constant 
(see menu 2.1.3.) 

0.8 ... 1.3 

U REF 

Internal voltage reference  
(see menu 1.1.3.) 

2.4. ... 2.6. 

T CST 

Time constant (see menu 2.2.4.) 

5 s 

TEST AMP 

Test value of the amplifier control 
loop (see menu 1.1.4.) 

75 ... 95 

FM 

Frequency of the magnetic field  
(see menu 1.1.1.) 

10 ... 14.5 Hz 

DIAMETER 

Diameter of pipeline  
(see menu 2.4.6) 

 50 mm 

Summary of Contents for DWM 2000 D

Page 1: ...DWM 2000 D DWM 2000 Electromagnetic Flowmeter with LCD Indicator Operating instructions...

Page 2: ...mable parameters 4 2 1 1 Flow calibration 4 2 1 2 Current output adjustment 4 2 1 3 Time constant 4 2 2 Electronics module checks 5 2 3 Programming structure Software n 1 02 5 2 3 1 User interface but...

Page 3: ...Operating Instructions DWM 2000 D 3 1 Display data in operating mode...

Page 4: ...a Vm with Va actual velocity Vm measured velocity reference value 2 1 2 Current output adjustment The minimum value i0 and the maximum value i100 of the current output at normal operating conditions c...

Page 5: ...and the corrective actions to be undertaken Call up the error messages menu 1 2 2 and note the last one Refer to the error message list Replace the electronics module in case of fatal error deactivate...

Page 6: ...6 Operating Instructions DWM 2000 D 2 3 2 Menu navigation Go through the steps given in the illustrations that follow to get to the required menu...

Page 7: ...Operating Instructions DWM 2000 D 7...

Page 8: ...eset the instrument to zero Refer to the installation manual to use the Zero Cal button on the DWM 2000 electronics block correctly 1 2 0 DIAGNOSTIC Sub menu 1 2 0 Diagnostic 1 2 1 ALARM MODE Deactiva...

Page 9: ...orised in the EEPROM see section 2 4 Parameters Stored in the EEPROM Read only 2 4 4 FILTER Activation of an electronic filter for noisy applications foam solid contents Select YES or NO default setti...

Page 10: ...test level CUSTOM CORRECTION YES NO Indication of activation of low velocity linearization YES EP CHECKS Result of the data check in EPROM see menu 1 1 6 Value 22309 With soft V1 02 FS Display of the...

Page 11: ...or 1 m s Replace the electronics module ZERO KEY Dysfunction of ZERO key short circuit Replace the electronics module EP CHECK Data loss in EPROM software Replace the electronics module EEP CHECK Data...

Page 12: ...e 7 0 095 911 7165 Fax 7 0 095 742 8873 krohne dol ru India Krohne Marshall Ltd A 34 35 M I D C Industrial Area H Block Pimpri Poona 411018 Phone 91 0 202 744 2020 Fax 91 0 202 744 2020 pcu vsnl net I...

Reviews: