KRIWAN INT280-60 Quick Start Manual Download Page 3

KRIWAN Industrie-Elektronik GmbH · Allmand 11 · D-74670 Forchtenberg · phone (+49) 7947 822 0 · fax (+49) 7947 1288 · e-mail: info@kriwan.com · home: www.kriwan.com

Technical changes reserved

Pr

oduct gr

oup S

Edition: 710.00566.0

INT280-60

®

INT280-60

®

Technical specifications

Order data

Supply voltage
- 52 S 581 S026

AC 50/60Hz 230V ±10% 25VA

- 31 S 581 S026

AC 50/60Hz 24V ±10% 25VA

Permissible ambient temperature 
T

A

-30...+60°C

Permitted rel. humidity

10-95% r. h. no condensation

Medium temperature

-30...+100°C

Operating pressure

-1...+60bar

Test pressure

90bar

Differential pressure

1-25bar

Relay
- Contact

AC 240V 2.5A C300 
at least AC/DC 24V 20mA

- Mechanical service life

Approx. 1 million switching cycles

Protection class acc. to EN 60529 IP65
Supply
- electrical

Push-in spring terminals
0.25-0.75mm

² 

- Oil

7/16"-20 UNF

Mounting

Flange (3-/4-hole)

Housing material

Aluminium
PA glass-fibre-reinforced

Flow rate through valve

1l/min (water 20°C, 1bar 
differential pressure)

Permitted oils

Mineral and ester oils without 
additives

Permitted refrigerants

Non-corrosive, non-flammable

Mounting position

Horizontal (rotatable by 180°), ±2°

Dimensions

See dimensions in mm

Weight

Approx. 1300g

Check base

EN 61000-6-2, EN 61000-6-3
EN 61010-1
Overvoltage category II
Pollution level 2

For additional technical details, see table of parameters

INT280-60 (AC 24V)

31 S 581 S026

INT280-60 (AC 230V)

52 S 581 S026

Accessories and application 
information

see www.kriwan.com

Table of parameters

Parameter name

Unit

Default

Filling cycle type

Fixed

Filling time

s

10

Break time

%

20

Overfill time

s

0

Oil deficiency alarm delay

s

120

Soiling stage 2

Warning

Reviews: