Operating Instructions
OIO 15en Edition 2020-05
30
Mounting instructions
Maintenance
Removing seals and bearings
Prerequisites:
□
Flowmeter removed from system
□
Sensors removed
1. Remove the socket screws
915.1*
, then
remove the end cover
070.1*
and O-ring
739.1
.
2. Pull the screw set with the ball bearings and
distance sleeve
039
out of the measuring
housing
128
, paying attention to the
configuration for later assembly. Remove the
distance sleeve.
3. Pull the ball bearings
817.1 + 817.3
and
817.2 + 817.4
off the measuring spindle
(large and small), using the pull-off device to
do so.
4. Remove the socket screws
915.2*
, then
remove the end cover
070.2*
and O-ring
739.2
.
5. Clean all the parts using a suitable detergent.
Installing seals and bearings
Prerequisites:
□
Replacement parts available
1. Insert the O-ring
739.2
into the measuring
housing
128
. Place on the end cover
070.2*
,
tighten the socket screws
915.2*
with torque,
2. Press the ball bearings
817.1 + 817.3
and
817.2 + 817.4
onto the measuring screws.
Notice
: Press on only over the inner ring!
3. Push the screw set into the measuring
housing
128
, paying attention to the
configuration. Slide the distance sleeve
039
flush into the hole of the measuring screw
small
672.2
.
4. Insert the O-ring
739.1
into the measuring
housing, place on the end cover
070.1*
.
Tighten the socket screws
915.1*
with torque,
A
817.2