16
e) Press “ENTER” key to save the current
calibration value.
f) Press “T1/T2/T1-T2” key (or “TC TYPE” key) to
switch to T1 measurement channel.
g) Input
0μV
in T1 measurement channel
h) After stabilizing, press “ENTER” key to save the
current calibration value.
i) Input 40.000mV in
T1 measurement channel
j) After stabilizing, press “ENTER” key to save the
current calibration value.
k) Press “T1/T2/T1-T2” key to switch to T2
measurement channel.
l) Input
0μV
in T2 measurement channel
m) After stabilizing, press “ENTER” key to save
the current calibration value.
n) Input 40.000mV in
T2 measurement channel
o) After stabilizing, press “ENTER” key to save the
current calibration value.
p) Power off and restart to complete calibration.
Note: Make sure your standard source is
accurate, then make calibration. The user is
responsible for using an accurate calibration
source. If the meter is inaccurate due using an
inaccurate calibration source, the user bears
the liability.
15
6.1Data Storage
Data is stored with two methods: manual storage
and automatic storage
1. Manual storage is in the normal measurement
mode, and data storage interval (INTER
V
AL) is
set to “00:00”. Press the “MEM” key to store an
item of data. The storage location is the smallest
number in the unused storage space. When
storage space is full of data, “FULL” character
will display for two seconds.
2. Automatic storage is in the normal measurement
mode, and data storage interval (INTER
V
AL) is
not set to “00:00”. Press the “MEM” key to start
storing data. The storage location is the smallest
number in the unused storage space. At the
same time, “ ” symbol will flash on the screen.
Press the “MEM” key again to stop storing data.
Each data storage time interval is the time set in
the data storage interval setting (INTER
V
AL).
The minimum time interval which can be set is 1
second, the maximum time interval is 59 minutes
and 59 seconds. When storage space is full of
data, “FULL” character will display for two
seconds, and stop storing data.
6.2 Data Reading
In the normal measurement mode, press the “READ”
key to enter data reading mode. At the same time,
“
” symbol will flash on the screen. Press
to
read the next data forward or backward. Hold
▼
to quickly move forward or backward in the stored
data. At the same time, “
” and number will shown
on the auxiliary display. If there is no available
storage space, “NULL” character will display for
two seconds.
REC
MEM
READ
6. Data Storage
(only KPS-TM340)