4. Setting Meter
4.1 SETUP Option
Press the SETUP key to enter setup mode. “ ”
symbol will flash on the screen, Press it again to
switch setting states sequentally. Save all previous
settings before exiting. If the meter is turned off in
the process of setting, setting changes will not be
saved. Changed settings will take effect
immediately after exiting setup mode.
Loop sequence: data storage interval setting
(INTER
V
AL) - thermocouple type setting (TYPE)
- OFFSET
(T1) setting - OFFSET (T2) setting -
automatic power-off time setting - system time
setting (S-T) - power frequency setting (LinE) -
normal temperature compensation (NTC) switch
setting - auto-calibration switch setting (CAL) -
data cleaning (CLr) - save setting and return to
normal measurement state (if self-calibration is
set to ON, return to the calibration state, then turn
the meter off and restart to return to normal
measurement state).
Note:
only
KPS-TM320
and
KPS-TM340
have OFFSET
(T2) option.
Only
KPS-TM340
have data storage
interval setting (INTER
V
AL) and data cleaning
(CLr) options.
4.2 SETUP Option Setting
1. Data storage interval setting (INTER
V
AL):
Press the SETUP
to enter the data storage
interval setting mode. “ ” will flash on the
screen. Use
to change automatic data
storage time interval. Hold
to quickly
increase or decrease setting time. Maximum
interval time is 59:59, minimum interval time
is 00:00 (00:00 enables automatic data storage
function; it only can be stored manually).
2.
Thermocouple type setting (TYPE)
Enter thermocouple type setting mode. Use
to set thermocouple types: K, J,
T
, E (
for KPS-TM340
R, S, N can be set, too. For
KPS-TM300
,
thermocouple types can be
switched directly with “TC TYPE” key)
3. OFFSET
(T1) setting
Users can adjust the thermometer displayed
value to compensate for a certain kind of
thermocouple error. See “Adjust the temperature
sensor error with deviation value” section.
Allowable adjustment range is ±6°C.
After entering OFFSET (T1) setting state, the
offset value setting can be changed with .
Note: When this offset value is no longer
required, please restore this offset value to 0.0.
Changing the thermocouple type will
automatically restore offset value to 0.0.