CAUTION
When using test leads or probes, keep your fingers behind the finger guards.
Use caution with voltages above 30 Vac rms, 42 Vac peak, or 60 Vdc. These voltages
pose a shock hazard.
If the test lead is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
To reduce the risk of fire or electric shock, do not use this product around explosive
gas or in damp locations.
DO NOT USE the test leads when the internal white insulation layer is exposed.
DO NOT USE the test leads above maximum ratings of CAT. environment, voltage and
current, that are indicated on the probe and the probe tip guard cap.
DO NOT USE the test leads without the probe tip guard cap in CAT III and CAT IV
environments.
DO NOT USE the test leads to measure over 1000V that is connected to MAINS
directly.
Maintenance
Do not attempt to repair this test lead set. It contains no user-serviceable parts. Repair
or servicing should only be performed by qualified personnel.
Cleaning
Clean the test lead with a water and mild detergent. DO NOT use abrasives or
solvents and DO NOT IMMERSE in liquid.
Specification
Input Impedance:
0ȍ
Overvoltage Category:
CAT 0 1500V AC, 2000V DC
CAT II 1000V AC, 1500V DC
CAT III 1000V, CAT IV 600V.
Pollution Degree 2
Exposed probe tip length :
19 mm to 4 mm (0.75 inch to 0.16 inch)
Environmental ratings :
-10°C to 45°C (-4°F to 113°F), 80% R.H.
Altitude :
2000 m (6,562 ft)
Safety Standard :
EN61010-031
CAT
Application field
The building installation.
The circuits directly connected to Low-voltage installation.
The source of the Low-voltage installation.
Ⅱ
Circuits that are not directly connected to Mains
0
Ⅲ
Ⅳ
Symbols as marked on the test lead and Instruction card
!
Risk of electric shock
DC measurement
Earth ground
AC measurement
Both direct and alternating current
Conforms to EU directives
See instruction Card
~
~—
Equipment protected by double
or reinforced insulation