KPS-LX30LED • Digital light meter
ENG
Figure 5-1-4A
Figure 5-1-4B
Notice
• Before zero calibration, the opto-sensor must be covered with the opto-
sensor visor.
Tips
• In non-zero calibration mode, press and hold “HOLD/ZERO” key for a short
time to enter the data hold mode.
• Zero-calibration can be made in any mode.
5.5 Source selection mode
• Hold the “MAX/MIN/LS” button for one second and the LS value will flash on
the LCD display (as shown in figure 5-1-5). The meter is now in source se
-
lection mode. Press “REL/PEAK” or “HOLD/ZERO” to change the light sou
-
rce (L0-L9), long press to accelerate change. Press “RAN” again to switch
back to the previous step.
• Hold the “MAX/MIN/LS” button to save the light source setting and return to
normal mode.
• Light source default values:
-
L0 - Standard lighting: 1.000
-
L1 - LED daytime light: 0.990
-
L2 - LED red light: 0.516
-
L3 - LED amber (yellow) light: 0.815
-
L4 - LED green light: 1.216
-
L5 - LED blue light: 1.475
-
L6 - LED purple light: 1.148