45
KPS-PA70 • Digital clamp meter
ENG
•
Insert black probe to COM jack, insert red probe to INPUT jack.
•
Measuring switch is placed to position .
•
Press “FUNC” key to switch to circuit continuity measuring state.
•
Connect the probe to the both ends of circuit to be tested for measurement.
• If the resistance of circuit to be measured is less than 50Ω, the meter’s built-
in buzzer may sound.
•
Read the circuit resistance value on the LCD.
NOTE:
If the probe is open or circuits resistance to be tested is more than 600Ω, the
display will show “0L”.
4.16 Capacitance Measurement
WARNING
Electric shock hazard. To avoid electric shock, before measuring ca-
pacitance, discharge capacitance completely
•
Insert black probe to COM jack, insert red probe to INPUT jack.
•
Measuring switch is placed to position .
•
After discharging capacitance completely, connect the probe to the both
ends of capacitor to be tested for measurement.
•
Read the capacitance on the LCD.
NOTE:
To improve the accuracy below 10nF measuring value, subtract the distributed
capacitance of meter and cable.
4.17 Surge Current Measurement
WARNING
Electric shock hazard. Remove the probe from the meter before mea-
suring with current clamp.
•
1) Measuring switch is placed to position A.
• 2) Hold the trigger, open clamp head, clip one lead of measurement circuit
to be tested in the clamp.
• 3) Press “REL/INRUSH” button more than 2 sec. to enter surge current mea
-
surement mode. The LCD will show “- - - -”, until motor start is detected. The
Summary of Contents for 602150001
Page 68: ...68 KPS PA70 Pince ampérométique numérique FRA ...
Page 69: ...69 KPS PA70 Pince ampérométique numérique FRA ...
Page 70: ...70 KPS PA70 Pince ampérométique numérique FRA ...
Page 71: ...71 KPS PA70 Pince ampérométique numérique FRA ...
Page 72: ...72 KPS PA70 Pince ampérométique numérique FRA ...
Page 73: ...73 KPS PA70 Pince ampérométique numérique FRA ...