8.Concentration Datalog Output Instruction
GT -1000 series gas detector allow user output history datalog via the charge port,there are several steps to output datalog.
1) Install concentration datalog output software in host computer
2) Boot up the detector and wait for 30s to preheat the instrument and connect to the computer
3) Run the concentration datalog output software in the host computer,make sure the detector is connected to the host
computer,meanwhile at the lower left interface of the software will show connection port standby(defaulted baud rate is
115200,do not change)
4) Enter the history data interface to output datalog,there will be menu prompts at the both software and detector interface.
5) User can preset detector’s address in concentration datalog output software(this address need to corresponded to the
parameter in “address setting”of the detector)
、
Channel(Corresponded to the gas channel of the detector)
、
start time and
finish time(datalog storage period should be within the start and finish time)
6) Detector allows user to search concentration of certain channel of datalog storage period,datalog can be output as EXCEL
format by “datalog output” function.
NOTICE:
Connection cable is the USB cable connected to the adapter