21
Emission test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1
The nebulizer uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The nebulizer is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
6.2 Electromagnetic immunity
The device is intended for use in the electromagnetic
environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
IEC60601
TEST LEVEL
Compliance
level
Electromagnetic
environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative
humidity should be
at least 30%.