![Konsung 408D User Manual Download Page 8](http://html1.mh-extra.com/html/konsung/408d/408d_user-manual_1996188008.webp)
8
No misting or
low rate of
misting
1. Check that there is medication in the
Nebulizer cup.
2. Check the Nebulizer for damage.
3. Check the position of the orange jet
inside the Nebulizer.
4. Make sure that air tube and other
components are properly attached.
5. Check the air filter and replace if
necessary.
6. Check the air tube and replace if
necessary.
Replacement Parts
Replacements parts are available through your local retailer or they can be
ordered by mail. See separate replacement part list for details.
Product Specification
Name and Model Number:
Air Compressor Nebulizer, # 408D,
Power:
AC 220V
±
22V, 50Hz
±
1Hz
AC 110V
±
11V, 60Hz
±
1Hz
Current
0.8A
Power Consumption
80W
Medicine Capacity
5mL
Sound Level
≤
58 dB
(
A
)
Compressor Pressure Range
210kPa to 250 kPa
Operating Pressure Range
50 kPa -100 kPa
Nebulization Rate
≥
0.2 mL/min
Operating Temperature Range 10
℃
-40
℃
Operating Humidity Range
10% RH-90% RH
Storage Temperature Range
-20.0
℃
-70.0
℃
Dimensions
220*140*245(mm
3)
Weight
1.31KGS
Standard Accessories
Nebulizer, Air Tube, Angled Mouthpiece, Air Filters,
Optional Accessories
Tee Adapter set, Baby Pacifier Inhalator set, Child
Mask, Adult Mask, Child’ s Mask, Air Filters
Specification are subject to change without notice.
Warning:
There must be Protection against electric shock on Class II Equipment and
protection on Type B applied parts, such as mouthpiece, baby pacifier
inhaler & mask.