Measuring method
Triangulation by light sectioning method
Light source
Semiconductor laser, Wavelength: 660 nm
Laser class
Class 2 (IEC 60825-1 Edition 2)
Number of pixels taken
1.31 megapixels (1280 x 1024)
Measurement distance
450 to 800 mm
Receiving lens
(interchangeable)
TELE, WIDE
Measure-
ment
range
(mm)
TELE
450 mm
TELE
800 mm
WIDE
450 mm
WIDE
800 mm
X × Y
79×99
141×176
150×188
267×334
Z
54
97
109
194
XY direction
measurement interval
0.08
0.14
0.16
0.28
Accuracy (Distance
between balls)*
1
±40μm
Precision (Z, σ)*
2
4μm
Auto-focus
Provided
Auto-exposure
Provided
Scan time
Approx. 2 sec. or more (1 scan)
Preview function
Approx. 0.4 sec./scan
Scanning ambient
lighting condition
500 lx or less
Output interface
USB 2.0 High Speed
Power
Included AC Adapter
Input voltage: 100 to 240 VAC (50/60 Hz), Rating: 1.4 A (at 100 VAC input)
Dimensions
295 (W) x 190 (H) x 200 (D) mm (Excluding grips and lens)
Weight
Approx. 6.7 kg
Operating tempera-
ture/humidity range
10 to 40°C; Relative humidity 65% or less (no condensation)
Storage tempera-
ture/humidity range
-10 to 50°C; Relative humidity 85% or less
(at 35°C, no condensation)
*1 When measuring the distance between balls for a ball bar (two balls) as defined in VDI/VDE 2634-2 under Konica
Minolta Sensing measurement conditions.
Konica Minolta Sensing measurement conditions: Temperature: 20±1°C; Lens: TELE; Measurement distance: 450
mm; Warm-up: 20 min.; Software used: Konica Minolta processing software; After calibration of the instrument;
Measurement target: Konica Minolta standard target (2 balls); Arrangement of measurement target: Konica Minolta
standard arrangement (10 positions inside measurement space); Uncertainty of standard target not included.
*2 Measurement conditions: Temperature: 20±1°C; Lens: TELE; Measurement distance: 450 mm; Warm-up: 20 min.; Mea
-
surement target: Konica Minolta standard planar chart; Software used: Konica Minolta processing software; 1 σ value
*3 Using Konica Minolta processing software
Specifications of 3D Data Processing Software RANGE VIEWER
<Major Functions>
Readable data
formats
Readable data formats
Konica Minolta proprietary formats: .rgv (Single-scan data group), .rvm
(more than one piece of data)
Data output
STL (binary), ASCII point group
Konica Minolta proprietary formats: .rgv, .rvm
Measurement
functions
Monitor image, Preview, AF/AE, Measurement
Control of rotating stage (Konica Minolta optional accessory)
E di t ing f unc-
tions
Data alignment, Data merging (integration), Point group deletion
Drawing
Point group shading
System configuration
Interchangeable
Objective Lens
(TELE, WIDE)
KONICA MINOLTA
RANGE7
AC Adapter
USB Cable
(5m) RA-A101
: Europe only (3m)
Marker Sheet Trial Pack
(Sheets T and W, 10 for each)
RANGE VIEWER
Personal
Computer
(commercial product)
Standard accessories
Optional accessories
3D Data Processing Software
Rotating Stage Kit (3 kg)
Rotating Stage Kit (20 kg)
Platform Set for Tripod
RA-A151
Measuring Stand
Platform Set for
Measuring Stand
RA-A201
Marker Sheet T
RA-A205
(50 markers x 10 sheets)
Marker Sheet W
RA-A206
(50 markers x 10 sheets)
Storage Case
RA-A209
Reverse engineering/
testing software
Calibration Sheet
Calibration Chart
RA-A105
RA-A107
Calibration Base
RA-A106
RA-A104
Dolly
Tripod
<Operating Environment>
OS
Windows® Vista 64-bit (x64)
CPU
Core2Duo, Xeon or higher
RAM
4 GB or more
Display
Graphics display capability of 1280 x 1024 or more
Graphics board
OpenGL compatible board (Konica Minolta verified compatible board recom
-
mended)
Interface
USB 2.0 port
Distance
Direction
SAFETY PRECAUTIONS
Read all safety and operating instructions before operating
the KONICA MINOLTA RANGE7.
Use only a power source of the specified rating.
Improper connection may cause a fire or electric shock.
Do not stare into the laser beam.
(MAX. 34mW 660nm , CLASS 2 (IEC) LASER PRODUCT)
34mW
http://www.konicaminolta.com/instruments/about/network
Addresses and telephone/fax numbers are subject to change without notice. For the latest contact information,
please refer to the KONICA MINOLTA SENSING Worldwide Offices web page .
Registration No. : YKA 0937154
Registration date : March 3, 1995
Registration No. : JQA-E-80027
Registration date : March 12, 1997
Osaka, Japan
EMail : [email protected]
Web : http://konicaminolta.com/instruments
Konica Minolta Sensing Americas, Inc.
New Jersey , U.S.A. Phone:
888-473-2656 (in USA), 201-236-4300 (outside USA)
Fax:
201-785-2482
EMail : [email protected]
Web : http://se.konicaminolta.us/3d
Konica Minolta Sensing Europe B.V.
German Office
München, Germany
Phone:
+49 (0) 89 630267-9700
Fax:
+49 (0) 89 630267-9799
EMail : info.
germany
@seu.konicaminolta.eu
Web : http://www.konicaminolta.eu
French Office
Roissy CDG, France
Phone :
+33(0)1 493-82519
Fax:
+33(0)1 493-84771
EMail : [email protected]
Web : http://www.konicaminolta.eu
UK Office
Milton Keynes, United Kingdom
Phone :
+44(0)1908 540-622
Fax:
+44(0)1908 540-629
EMail : [email protected]
Web : http://www.konicaminolta.eu
Italian Office
Milan, Italy
Phone :
+39 02 390111
Fax:
+39 02 39011.223
EMail : [email protected]
Web : http://www.konicaminolta.eu
Polish Office
Wroclaw, Poland
Phone :
+48 71 734 52 11
Fax:
+48 71 734 52 10
EMail : [email protected]
Web : http://www.konicaminolta.eu
Konica Minolta (CHINA) Investment Ltd.
SE Sales Division
Shanghai, China
Phone:
+86-021-5489 0202
Fax:
+86-021-5489 0005
Konica Minolta Sensing Singapore Pte Ltd.
Singapore
Phone:
+65 6563-5533
Fax:
+65 6560-9721