48
5�1
Maintenance and inspection items
This chapter describes the inspections and cleaning
required in order to maintain the use of this device in
an optimum condition
Quality Tool
HINT
•
"Simple Check QC for ImagePilot" is available for checking
of imaging quality. Please contact our customer service for
the detail.
•
For details, contact Konica Minolta technical representa-
tives.
5�1�1
Maintenance schedule
The maintenance and inspection items that the user
should perform are as follows.
IMPORTANT
•
To ensure optimum use of this device, be sure to perform
periodic maintenance
•
The following task intervals are estimates and vary accord-
ing to usage.
•
If image burn-in or defect pixels in the image is visible, con-
tact Konica Minolta technical representatives.
Maintenance task
Maintenance
interval
Checking and cleaning the surface of the
DR Detector
Weekly
Checking for external damage to the DR
Detector
Weekly
Cleaning the DR Detector and wired
connection connectors
Weekly
Cleaning the DC Power Cable and con-
nectors
Weekly
Cleaning the Battery Charger
Weekly
Full charge of the Battery
6 months
5�1�2
Cleaning
IMPORTANT
•
Be careful not to apply any cleaning chemical or liquid onto
the wired connection connectors, the Power SW, the Selec-
tion SW, and the LEDs.
•
Do not clean with sharp or hard metal objects. If you cannot
remove stains, contact Konica Minolta technical represen-
tatives.
•
Pay attention not to damage the wired connection connec-
tors or the battery connectors port of the DR Detector, the
Battery, and the Battery Charger.
Exterior
IMPORTANT
•
If you directly apply or spray antiseptic solution on the DR
Detector, the solution may enter the inside of the instru-
ment through exterior gaps, causing a failure.
•
Clean the exterior with a soft lint-free cloth damp-
ened with a small amount of antiseptic solution and
wrung well.
•
To clean the gaps on the exterior of the DR Detector,
remove dirt using a commercial plastic brush.
Wired connection connector
If foreign material has adhered to the wired connection
connectors of the DR Detector and DC Power Cable,
and/or the battery connector ports of the DR Detector,
the Battery, and the Battery Charger, remove it with a
commercial plastic brush.
DR Detector
Wired connection connector
Battery connector port
Summary of Contents for AeroDR SKR 4000
Page 1: ...EN 02...
Page 2: ......
Page 8: ...8...
Page 9: ...9 Chapter 1 Safety Precautions Warnings...
Page 22: ...22...
Page 23: ...23 Chapter 2 Product Overview...
Page 29: ...29 Chapter 3 General Operations...
Page 42: ...42...
Page 43: ...43 Chapter 4 Troubleshooting...
Page 46: ...46...
Page 47: ...47 Chapter 5 Maintenance...
Page 50: ...50...
Page 51: ...51 Chapter 6 Specifications...
Page 56: ...56...
Page 57: ...57...
Page 58: ...58...
Page 59: ......