K244XX Programmable Ashing Furnace
Operation and Instruction Manual
K244XX-Manual
-
28-
Program Settings
The program settings are used to select the mode of operation for the automatic ramp/soak programs. The
program Settings screen is accessed from the “Program” menu.
The
Single
mode selection button allows each control loop to run independently of the others. Programs are
then downloaded separately to each loop control. The user can then individually start and stop the program
in each loop control. Any combination of loop controls can then be operated in single set point (static) mode
or automatically via the program at any given time.
The
All
mode selection button links all loop controls into a master/slave program operation. While in single
set point (static) mode, each loop control operates independently from the others. However, when an
automatic program is started, the loop defined as the “Primary” loop becomes the master and all other loops
become set point slaves.
When the ramp/soak program is started in the primary loop, the display will automatically set the set point of
all other (slave) loops to match the set point of the primary loop. The user will be unable to manually change
the loop set points of the slave loops during program operation. However, any events tied to the slave loops
can be manually turned on and off during program operation. Only events tied to the primary loop will be
controlled via the automatic ramp/soak program.
The
Primary
loop selection is used to select which loop is to be used as the master loop when operating
under the “All” program mode.