Kodak V2CS Technical Data Manual Download Page 7

KODAK VISION2 Color Negative Control Strips, Technical Data and Crossover Information

7

Effects of Mechanical and Chemical 
Variations

Figures 4 through 13 illustrate some of the major 
photographic effects of mechanical and chemical variations 
on KODAK VISION2 Color Negative Control Strips 
exposed on 5217 Film compared to the former VISION 
Control Strips exposed on 5274 Film. Each plot shows the 
effect of a change in a process variable (horizontal axis), on 
the dye density of the processed film (vertical axis). These 
density plots are deviated against the standard level for each 
variable (e.g. standard level for the variable is represented by 
zero density).

The magnitude of the changes shown in these plots should 

not be considered to be process control limits. Also, the data 
presented are qualitative, not quantitative. The plots were 
derived from experiments using small laboratory machines 
in which all constituents were held constant except the 
variable being studied. Hence, the figures should be used 
only as trend charts and guides. If two or more process 
variable are changed, the resulting photographic effect 
illustrated may not be additive. Interactions can occur that 
produce effects other than those predicted by addition. The 
plots in this publication are representative only; they do not 
contain all possible solution problems. Most of the important 
photographic effects take place in the developer.

Summary of Contents for V2CS

Page 1: ...Eastman Kodak Company 2004 V2CS August 2004 KODAK VISION2 ColorNegative Control Strips Process ECN 2 Technical Data and Crossover Information...

Page 2: ...274 5 ECN 2 Control Strip Exposed on KODAK VISION2 200T Color Negative Film 5217 6 Effects of Mechanical and Chemical Variations 7 Developer Mechanical Factors 8 Effects of Time and Temperature Variat...

Page 3: ...ulsion is being used for process control The new format is exposed on KODAK VISION2 200T Color Negative Film 5217 while the former ECN control strip was exposed on KODAK VISION 200T Color Negative Fil...

Page 4: ...Negative Film 5217 Physically the center to center step distance has been maintained on the new control strips no problems should be encountered with existing automatic densitometers Once the transiti...

Page 5: ...Data and Crossover Information 5 Figure 2 ECN 2 Control Strip Exposed on KODAK VISION 200T Color Negative Film 5274 0 0 1 0 2 0 3 0 4 0 3 0 2 0 1 0 0 0 1 0 2 0 Densitometry Exposure Process ECN 2 1 5...

Page 6: ...l Data and Crossover Information Figure 3 ECN 2 Control Strip Exposed on KODAK VISION2 200T Color Negative Film 5217 Exposure 1 50 sec Process ECN 2 Densitometry Status M 0 0 1 0 2 0 3 0 4 0 3 0 2 0 1...

Page 7: ...level for the variable is represented by zero density The magnitude of the changes shown in these plots should not be considered to be process control limits Also the data presented are qualitative no...

Page 8: ...320 F Time 5217 20 10 15 25 05 0 05 15 20 25 10 20 10 15 05 0 10 05 AIM STRIP CODE NO REFERENCE VALUES RED B W BLUE GREEN B W BLUE GREEN RED 240 220 340 15 10 20 0 05 05 300 240 10 0 05 220 05 10 10 3...

Page 9: ...H 5217 20 10 15 25 05 05 0 10 20 15 25 15 20 05 10 0 VALUES REFERENCE CODE NO AIM STRIP GREEN BLUE RED B W GREEN BLUE RED 10 20 15 10 0 05 05 10 10 05 0 05 10 15 10 05 10 45 10 25 10 15 10 05 10 35 10...

Page 10: ...2 1 4 0 8 1 6 1 0 1 2 1 4 0 8 NaBr 5217 20 15 10 25 05 05 0 10 20 15 25 15 20 05 10 0 VALUES CODE NO REFERENCE STRIP AIM BLUE GREEN RED B W RED BLUE GREEN 10 20 15 10 0 05 05 10 10 05 0 05 g L 10 B W...

Page 11: ...Na 5217 20 10 15 25 05 05 0 10 20 15 25 15 20 05 10 0 VALUES REFERENCE CODE NO AIM STRIP GREEN BLUE RED B W GREEN BLUE RED 10 20 15 10 0 05 05 10 10 05 0 05 3 0 2 0 1 5 1 0 2 5 10 B W GREEN RED B W BL...

Page 12: ...0 150 0 50 0 0 0 100 0 mL L 200 0 50 0 100 0 mL L 150 0 0 0 200 0 100 0 150 0 PB C1 5217 20 10 15 25 05 05 0 10 20 15 25 15 20 05 10 0 REFERENCE CODE NO STRIP VALUES AIM GREEN BLUE RED B W GREEN BLUE...

Page 13: ...5 10 B W BLUE GREEN g L Substitute CD 2 for CD 3 5274 05 BLUE GREEN RED B W BLUE B W GREEN RED B W 10 1 5 0 0 0 5 g L 1 0 2 0 0 0 0 5 05 05 0 10 10 05 05 0 10 20 15 10 GREEN BLUE RED B W GREEN BLUE RE...

Page 14: ...cess ECN 2 Developer HD LD D Min LD 0000 HD MD 10 0 05 RED 05 10 B W BLUE GREEN mL L 5274 BLUE B W GREEN RED BLUE GREEN B W RED mL L 4 0 10 10 05 05 0 10 0 0 2 0 05 05 0 20 10 15 6 0 RED BLUE GREEN B...

Page 15: ...l Rinse mL L 300 100 200 0 mL L F 34a Fixer 5217 20 10 15 25 05 0 05 15 20 25 10 20 10 15 05 0 10 05 AIM STRIP CODE NO REFERENCE VALUES RED B W BLUE GREEN B W BLUE GREEN RED 0 75 15 10 20 0 05 05 0 25...

Page 16: ...5 RED 05 10 B W BLUE GREEN 5274 2 5 mg L 05 RED GREEN BLUE B W BLUE B W RED GREEN B W 10 5 0 mg L 0 0 2 5 7 5 0 0 05 05 0 10 10 05 05 0 10 15 20 10 RED BLUE GREEN B W RED BLUE GREEN AIM STRIP VALUES C...

Page 17: ...6 25 6 15 6 5 6 45 6 35 6 25 6 HD LD D min LD MD HD 0000 2 Na S Bacterial 5217 20 10 15 25 05 05 0 10 20 15 25 15 20 05 10 0 REFERENCE CODE NO VALUES STRIP AIM GREEN BLUE RED B W GREEN BLUE RED 10 20...

Page 18: ...y the new batch trends as compared to the previous batch this is a normal batch to batch variation and is to be expected If you are unable to maintain a plot within your action limits with the new con...

Page 19: ...ord Form H 24F Control Strip Crossover Process ECN 2 DATES GREEN B W BLUE GREEN GREEN B W RED RED BLUE B W BLUE RED 05 10 0 CB control 0 08 CB action 0 06 CB control 0 09 CB action 0 07 10 10 05 10 0...

Page 20: ...DAK Publication No V2CS MORE INFORMATION Outside the United States and Canada please contact your Kodak representative You can also visit our web site at www kodak com go motion for further informatio...