8.1 Module
replacement
MODCAP modules are delivered with DIP-SWITCHES 2, 3 and 4 adjusted with its own
probe. In case of replacement, following procedure must be done with the probe free of
product:
1. With DIP-SWITCH 1 to “MAX” and 2, 3 and 4 to “MIN”, turn the multiturn
potentiometer totally to clockwise (20 turns). RED LED will switch ON.
2. In case that RED LED doesn’t lights, move DIP-SWITCH 4 (drawing 1) to
ON. If RED LED still remains OFF, do same with DIP SWITCH 3 (drawing 2)
and finally with DIP-SWITCH 2 (drawing 3) until RED LED finally lights.
Drawing 1
Drawing 2
Drawing 3
3. Once DIP-SWITCHES are correctly settled the SENSITIVITY ADJUSTMENT
must be done (chapter 8).