8. Adjustment
After installation and connection verification, we proceed to AJUST SENSITIVITY to fit
the NSC with the tank and the product to measure.
Check that product doesn’t reach the probe. Power supply GREEN LED must be ON
1. Put DIP-SWITCH 1 to “MAX” position.
2. Turn the “SENSITIVITY ADJUSTMENT screw” clockwise until RED LED
lights.