NMC
DT0305
Page
7
8. To avoid mistakes, asks again if you are sure that
you want to calibrate the unit. Press ( OK ) another
time to enter.
9. Pressing (+) we enter in High level calibration.
10. If when you adjust the high level, the liquid in the
tank is 80 % of the capacitance, we will indicate it in
the display using the (+) and (-) keys. Confirm with
(OK).
11. The microprocessor makes the operations to
calibrate the maximum level of the unit. This
operation takes some seconds.
12. The display shows this message to indicate NMC
has finished the calibration internal process.
When the liquid level will be in the minimum point, it
will indicate 000.0 % and when the liquid level will
be in the maximum, 100.0 %.
13. Automatically, the display shows the main menu.
Pressing ESC the unit returns to read state.
Pressing (+) we can enter in output adjustment
menu and simulation mode.
9.2 Output
Adjustment
1. Pressing any key, you enter in the main menu.
Press (+) to enter in Output.
2. Pressing (+), we enter in Output adjustment menu.
CALIBRATION
¿SURE?
Yes
OK No
ESC
CALIBRATION
Low Level
OK
High Level
+
HIGH LEVEL ADJUST
Enter the actual level
In probe
+ and - :080.0 %
OK
Calibrating
PROBE
For low level
…. WAIT ….
*** PROBE
CALIBRATED ***
Calib.Probe
OK
Output
+
Language
-
Back
ESC
Calib.Probe
OK
Output
+
Language
-
Back
ESC
OUTPUT
Output SIM
-
Output Adj
+
Back
ESC
Summary of Contents for NMC Series
Page 1: ...Operating instructions for Capacitive level transmitter Model NMC 2 Wire 4 to 20mA...
Page 15: ...NMC DT0305 Page 15 16 ATEX Certified...
Page 16: ...NMC P ge 16 DT0305...
Page 17: ...NMC DT0305 Page 17...
Page 18: ...NMC P ge 18 DT0305...
Page 19: ...NMC DT0305 Page 19...
Page 20: ...NMC P ge 20 DT0305 17 Models...
Page 22: ...NMC P ge 22 DT0305 19 Notes...
Page 23: ...NMC DT0305 Page 23...