NMC
Páge 6
DT0305
9. Programming
9.1 Calibrating
probe
1. Pressing any key, you enter in the main menú.
Press (OK) to enter in calibration probe.
2. To avoid mistakes, asks again if you are sure that
you want to calibrate the unit. Press (OK) another
time to enter.
3. If the level in the tank is low, we will choose the
(OK) option to adjust the unit with the lowest level
4. If, for example, when you adjust the low level, the
level of the tank is the 10 % of the capacitance, we
will indicate it in the display using the ( + ) and ( - )
keys. Confirm with ( OK ).
5. In this moment the microprocessor makes the
operations to choose the best measurement range.
This operations takes some seconds.
6. When this operation is finished, the display shows
the message “PROBE CALIBRATED”.
7. Automatically the display shows the main menu.
Now need calibrate the High Level, then pressing
(OK) another time to enter in the calibration probe
menu.
Calib.Probe
OK
Output
+
Language
-
Back
ESC
CALIBRATION
¿SURE?
Yes
OK No
ESC
CALIBRATION
Low Level
OK
High Level
+
LOW LEVEL ADJUST
Enter the actual level in probe.
+And-: 010.0 %
OK
Calibrating
PROBE
For low level
….WAIT….
***PROBE
CALIBRATED**
Calib.Probe
OK
Output
+
Language
-
Back
ESC
Summary of Contents for NMC Series
Page 1: ...Operating instructions for Capacitive level transmitter Model NMC 2 Wire 4 to 20mA...
Page 15: ...NMC DT0305 Page 15 16 ATEX Certified...
Page 16: ...NMC P ge 16 DT0305...
Page 17: ...NMC DT0305 Page 17...
Page 18: ...NMC P ge 18 DT0305...
Page 19: ...NMC DT0305 Page 19...
Page 20: ...NMC P ge 20 DT0305 17 Models...
Page 22: ...NMC P ge 22 DT0305 19 Notes...
Page 23: ...NMC DT0305 Page 23...