Configuration Set NGM-HART
page 12
NGM-HART K05/0922
5.3 Response Time
BASIC CONFIGURATION
• Read out actual response time, by clicking on I9.
Field G9 is showing the actual response time multiplied with 0,1ms.
• Change actual response time within a range of 2 … 100 (0,2 … 10 sec) in field
G8 and clicking on I8 “set response time”.
Use high response times for storage tanks with slow level movements. Use low response
times for buffer and process tanks.
Verify changes by clicking again on I9 “get response time”