Kobold KME Operating Instructions Manual Download Page 22

22

Operating instructions for KME flow meter for compressed air and gases

8  technical data

Measured values

Flow

Measurands  

   

   

m

3

/h, m

3

/min, l/min, l/s, kg/h, kg/min, m/s, SCFM, ft/min, °C, °F

Standard conditions (factory setting) 

1013.25 mbar 

(14.7 psi)

, 0 °C 

(32 °F) 

(configurable)

Measuring range 

in air

  

 

     

DN15:   0.2...76.3  Nm

3

/h     

(0.12...44.88 SCFM)

 

    

 

 

 

  DN20: 

 

0.4...135.6 

Nm

3

/h  

(0.24...79.77 SCFM)

 

    

 

 

 

  DN25: 

 

0.6...212 

Nm

3

/h       

(0.36...124.71 SCFM)

    

 

 

 

 

DN32:   0,9...347,4 Nm

3

/h  

(0.52...202.06 SCFM)

 

    

 

 

 

  DN40: 

 

1,4...542,8 

Nm

3

/h  

(0.81...315.71 SCFM)

 

    

 

 

 

  DN50: 

 

2,2...848,2 

Nm

3

/h  

(1.22...493.35 SCFM)

Accuracy

1)

 

in air

 

at 7 bar (102 psi) (abs) and 23 °C (73 °F) 

± (3 % of measured value + 0.3 % of full scale)

Temperature coefficient   

     

± 0.25 % of the measured value / °C deviating from 23 °C 

(73 °F)

Pressure coefficient

2)

  

 

     

+ 0.5 % of the measured value / bar deviating from 7 bar 

(102 psi)

Response time t

90

   

 

     

< 2 sec

Measuring rate 

 

 

     

0.1 sec

Temperature

Measuring range 

 

 

     

-20...60 °C 

(-4...140 °F)

Accuracy 

at 20 °C (68 °F) and flow >0.5 Nm/s

  

 

± 0.7 °C 

(1.26 °F)

Outputs

Analogue output (scalable) 

     

0 - 20 mA / 4 - 20 mA 

,    R

L

<500 Ohm

Switch output  

 

 

     

DC PNP, max. 100 mA, V

drop

 <2.5 V, 10 kOhm pull-down

 

 

 

 

 

     

Configurable: N/C or N/O, hysteresis, window

Pulse output   

 

 

     

Consumption volume meter, pulse duration 0.02...2 sec.

Bus interface  

 

 

     

Modbus RTU (max. 32 bus devices) or 

    

 

 

 

  M-BUS 

(Meter-Bus)

Configuration interface   

     

USB

General

Supply voltage 

 

 

     

18 - 30 V DC

Current consumption (max.)       

 

with display   

 

     

I

max

 

120 mA    (P

max

 

≤2,5 W)

 

without display 

 

     

I

max

 

60 mA  

 (P

max

 

≤1,6 W)

Operating pressure (max.) 

     

16 bar 

(232 psi)

 / PN16

Ambient temperature  

 

with display   

 

     

0...50 °C 

(32...122 °F)

 

 

without display 

 

     

-20...60 °C 

(-4...140 °F)

Medium and storage temperature   

-20...60 °C 

(-4...140 °F)

Humidity 

 

 

 

     

0...100 % rH, non-condensing

Medium 

 

 

 

     

Compressed air, nitrogen, oxygen, CO

2

, argon

Electrical connection  

 

     

M12x1 4 pin plug

Electromagnetic  compatibility       

EN61326-1 

   EN61326-2-3

    

 

 

 

  Industrial 

environment

Material  

 

Enclosure     

   

Polycarbonate

 

Probe tube   

 

     

Stainless steel 1.4404

 

Probe head / sensor       

Stainless steel 1.4404 / glass

 

Measuring block   

     

Aluminium anodised or stainless steel 1.4404

Enclosure protection class 

     

IP65

1)  The tolerance specifications include the uncertainty of the factory calibration with a coverage factor k=2 (2 x standard deviation). The tolerance was calculat-

ed in accordance with EA-4/02 following the GUM (Guide to the Expression of Uncertainty in Measurement).

2)  The flow meter was adjusted at 7 bar (102 psi) (abs). At an operating pressure other than 7 bar (102 psi) (abs), the error can be corrected by entering the 

actual system pressure (display menu or configurator software). 

Summary of Contents for KME

Page 1: ...BA_KME_e V1 3 Technical data subject to change 194573 Operating instructions KME Flow meter for compressed air and gases...

Page 2: ...high frequency energy If it is not installed and used in accordance with the operating instructions it may cause electromagnetic interference to radio communica tions However there is no guarantee tha...

Page 3: ...ure 10 3 3 Installation position 10 3 4 Inlet and outlet measurement path 11 3 5 Installation of the gauge mounting block 12 3 5 1 Operating the pipeline without transmitter 12 3 6 Mounting the transm...

Page 4: ...h optimal performance of the device 1 2 Safety instructions 1 2 1 Intended use The flow meter device is dedicated to the measurement of compressed air non corrosive and non flammable gases in pipeline...

Page 5: ...rized staff before putting the system into operation Do not install or start start up a device supposed to be faulty Make sure that such devices are not accidentally used by marking them clearly as fa...

Page 6: ...umption metering The KME features an integrated consumption meter The consumption volume can be shown on the display and is retained even if the supply voltage is interrupted Transmitter Display optio...

Page 7: ...r Software 2 2 Functions 2 2 1 Analogue output OUT 1 The analogue current output factory setting 4 20 mA or 0 20 mA is used for the actual flow or temperature measured values The analogue output is fr...

Page 8: ...mption is stored every minute and is retained even if the supply voltage is interrupted The totalized consumption can be shown on the display A reset of the consumption meter can be performed via disp...

Page 9: ...nting location for installing removing the transmitter unit of the E741 The ambient temperature shall not exceed the specified limits Consider also the possible heat radiation Air medium purity at the...

Page 10: ...y adjusted at 7 bar 102 psi absolute pressure For normal requirements pressure compensation is not necessary For best measure ment accuracy the actual working pressure can be set via display menu or v...

Page 11: ...complished by observing a minimum inlet and outlet path length which depends on both the nature of disturbance and the pipe diameter The flowmeter shall be located before valves or sliders With light...

Page 12: ...g 5 The sealing plug has a Tuflok coating http www bossard com and does not require any additional seal Mount the sealing plug with a torque of min 28 Nm Under normal operation with transmitter instal...

Page 13: ...ly fig 8 Fig 8 Tighten mounting screws 4 Electrical installation The electrical installation of the KME shall be performed by qualified staff only Observe all applicable national and international req...

Page 14: ...01 Version 09 Medium 9 compressed air XX Access number consecutive 00 Status 00 00 Signature Data record 1 Volume flow 05 DIF 32 bit real 3E VIF volume flow in m h XX XX XX XX Cur measured value Data...

Page 15: ...t float 30519 206 Standard volume flow Nm min 32 bit float 30521 208 Standard volume flow l min 32 bit float 30523 20A Standard volume flow l s 32 bit float 30525 20C Standard volume flow SCFM 32 bit...

Page 16: ...MAX value mass flow 16 bit integer 60008 7 Reset MAX value standard volume flow 16 bit integer 60009 8 Reset consumption meter 16 bit integer Data transmission Factory settings Adjustable values Baud...

Page 17: ...e on a personal computer 5 Display The LCD display optional shows the actual measured values and the overall consumption The complete KME setup and configuration can be performed with the control keys...

Page 18: ...dard flow Fig 16 Status display Fig 17 Measured value display Explanation of the symbols Output 1 set to analogue output Output 2 set to pulse output consumption Output 1 set to switching output statu...

Page 19: ...pe diameter Set the pipe diameter see 2 1 1 on page 7 Display Set the measuring mode and the display orientation Process parameter Set the operating pressure and the standard conditions see 2 2 4 page...

Page 20: ...LED lights up Cause The display or the communication with the display is faulty Consequence Analogue switch and pulse outputs operate normally Remedy Return the device to the manufacturer for service...

Page 21: ...or isopropanol Isopropanol is recommended for contamination with greases or oils Do not touch or rub the sensor element within the sensing head Allow the sensing head to dry free 7 3 Ordering guide f...

Page 22: ...er pulse duration 0 02 2 sec Bus interface Modbus RTU max 32 bus devices or M BUS Meter Bus Configuration interface USB General Supply voltage 18 30 V DC Current consumption max with display Imax 120...

Page 23: ...00 2500 250 Standard volume flow l s DN15 0 20 14 1 4 DN20 0 35 25 2 5 DN25 0 55 40 4 Standard volume flow SCFM DN15 0 44 30 3 0 1 0 05 DN20 0 76 53 5 3 0 15 0 07 DN25 0 117 88 8 8 0 2 0 1 Mass flow k...

Page 24: ...40 0 8300 5800 580 DN50 0 13300 10000 1000 Standard volume flow l s DN32 0 83 56 5 6 DN40 0 139 97 9 7 DN50 0 222 167 16 7 Standard volume flow SCFM DN32 0 176 117 11 7 0 55 0 25 DN40 0 294 200 20 0 9...

Page 25: ...ory use EMC requirements Part 1 General requirements eN 61326 2 3 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 2 3 Particular requirements Test configurat...

Page 26: ...26 Operating instructions for KME flow meter for compressed air and gases...

Page 27: ...27 Operating instructions for KME flow meter for compressed air and gases...

Page 28: ...www kobold com...

Reviews: