HND-T105
page 8
HND-T105 K04/1016
7.5.4
'Scal': Scale Correction
-2.000...2.000:
The scale of the measuring will be changed by this
factor to compensate for deviations in the temperature
probe or in the measuring device (factor is in %)
oFF:
Scale correction factor inactive (=0.000)
7.5.5
'Power.off': Selection of Power-Off Delay
1...120:
Power-off delay in minutes.
Device will be automatically switched off as soon as this
time has elapsed if no key is pressed/no interface
communication takes place
oFF:
Power-off function inactive (continuous operation, e.g.
mains operation)
7.5.6
'Out': Function of the Output
OFF:
No output function, lowest power consumption
SEr:
Output is serial interface
dAC:
Output is analogue output 0...1 V
7.5.7
'Address': Selection of Base Address when Output = Serial
Interface
01, 11, 21, ..., 91
:
Base address of device for interface communication.
7.5.8
'dAC.0Volt': Output Offset When Output = Analogue Output
-200.0...850.0 °C
respectively
-328.0...1562.0 °F
Enter desired temperature value at which the
analogue output potential should be 0 V
7.5.9
'dAC.1Volt': Output Scale When Output = Analogue Output
-200.0...850.0 °C
respectively
-328.0...1562.0 °F
Enter desired temperature value at which the
analogue output potential should be 1 V
Hint: The settings will be set to the settings ex works, if key ‘Set‘
and ‘Store‘ are pressed simultaneously for more than 2 seconds.