Kobold HND-P105 Operating Instructions Manual Download Page 6

HND-P105 

page 6 

 

HND-P105 K05/0117 

7.2 Connections  

1. 

Interface:

 Connect to optically isolated 

interface adapter (accessory: GRS 3100, 
GRS 3105, USB 3100 or USB 3100N) 

2. 

Connections

 for 

pressure sensors

 of the 

HND-PS-family 

3. The 

mains socket

 is located at the left side 

of the instrument 

7.3 Display 

 

 

7.4 Basic Operation 

When switching on the device displayed “Corr”, if a sensor with activated offset or 
scale correction is connected. 
 

 

 

Tare Function:  

By pressing `Tara`(key 3) the displays will be set to 0. All 

measuring from then on will be displayed relatively to the set tare 
values. When the tare function is activated, the arrow “Tara” 
appears in the display. To deactivate tare function press `Tara`for 
>2 seconds. 

Please note: Activating/deactivating tara clears 

the max-&min-memories.

 

Max Memory:

   Pressing ´max´ (key 2) shows the maximum of the measured 

values. Pressing it again hides it. To clear the max memory press 
key ´max´ for >2 seconds. 

Min Memory:

  Pressing ´min´ (key 5) shows the minimum of the measured 

 

values. Pressing it again hides it. To clear the min memory press 

 

key ´min´ for >2 seconds. 

Hold Function: 

By pressing ´Store/Quit´ (key 6) the last measuring value will be 

 

held in the secondary display. Pressing it again hides it. 

Summary of Contents for HND-P105

Page 1: ...Operating Instructions for Handheld Pressure Measuring Devices with External Pressure Sensors Model HND P105...

Page 2: ...uration 7 7 6 Measuring Of Water Level Display Unit m 8 7 7 Calibration Services 9 7 8 Pressure Connection To The Sensors 9 7 9 Error And System Messages 10 7 10 The Serial Interface 11 8 Maintenance...

Page 3: ...before shipping and sent out in perfect condition Should damage to a device be visible we recommend a thorough inspection of the delivery packaging In case of damage please inform your parcel service...

Page 4: ...upply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load voltage We therefore recommend using regulated voltage...

Page 5: ...make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be...

Page 6: ...uring from then on will be displayed relatively to the set tare values When the tare function is activated the arrow Tara appears in the display To deactivate tare function press Tara for 2 seconds Pl...

Page 7: ...on 7 5 1 Unit Choice Of The Display Unit Choose the desired display unit the referring unit is displayed by means of a functional arrow in the display The selection is permanently stored in the sensor...

Page 8: ...r deviations Input is in the display unit 7 5 6 SCAL Adjusting Sensor Scale The scale of the measuring can be influenced by this setting factor is in displayed value measured value 1 Scal 100 Standard...

Page 9: ...l dia of 4 mm to pressure port B Port A will not be used Pressure sensors HND PS01 PS02 and P03 allow for measurements of under pressure up to the entire over pressure measuring range by re plugging t...

Page 10: ...thout battery wrong voltage or polarity Check power supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufactu...

Page 11: ...nge unit 7 Read max value 204 Read display range decimal point 12 Read ID number 208 Read of channels 176 Read min measuring range 214 Read Scale adjustment 177 Read max measuring range 216 Read Offse...

Page 12: ...ower off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7mm high for measuring values and for min max memories hold functio...

Page 13: ...0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 350 0 mbar rel 0 1 mbar max 1 4 bar 0 70 C G male stainless steel HND PS09 0 1000 mbar abs 1 mbar max 4 bar abs HND PS10 0 3500 mbar abs max 14 bar abs...

Page 14: ...face converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with lo...

Page 15: ...ectrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic product...

Reviews: