Kobold HND-P105 Operating Instructions Manual Download Page 5

  

HND-P105 

HND-P105 K05/0117 

 

page 5 

7.  Operation / Configuration / Adjustments 

7.1 Safety Requirements 

This device has been designed and tested in accordance with the safety 
regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless 
the standard safety measures and special safety advises given in this manual will 
be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed 

if the device is not subjected to any other climatic conditions than those 
stated under 

9 Technical Information

2.  Device and sensors have to be handled with care (don’t throw, hit, etc.). 

Protect plugs and sockets from soiling. 

3. If the device is transported from a cold to a warm environment 

condensation may cause in a failure of the function. In such a case make 
sure the device temperature has adjusted to the ambient temperature 
before trying a new start-up. 

4.  If device is to be connected to other devices (e.g. via serial interface) the 

circuitry has to be designed most carefully. Internal connection in third 
party devices (e.g. connection GND and earth) may result in not-
permissible voltages impairing or destroying the device or another device 
connected.  

Warning:  If device is operated with a defective mains power supply 
(e.g. short circuit from mains voltage to output voltage) this may 
result in hazardous voltages at the device (e.g. at sensor socket or 
interface). 

5.  If there is a risk whatsoever involved in running it, the device has to be 

switched off immediately and to be marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 

  there is visible damage to the device 

  the device is not working as specified 

  the device has been stored under unsuitable conditions for a longer 

period of time. 

In case of doubt, please return device to manufacturer for repair or 
maintenance. 
 

Summary of Contents for HND-P105

Page 1: ...Operating Instructions for Handheld Pressure Measuring Devices with External Pressure Sensors Model HND P105...

Page 2: ...uration 7 7 6 Measuring Of Water Level Display Unit m 8 7 7 Calibration Services 9 7 8 Pressure Connection To The Sensors 9 7 9 Error And System Messages 10 7 10 The Serial Interface 11 8 Maintenance...

Page 3: ...before shipping and sent out in perfect condition Should damage to a device be visible we recommend a thorough inspection of the delivery packaging In case of damage please inform your parcel service...

Page 4: ...upply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load voltage We therefore recommend using regulated voltage...

Page 5: ...make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be...

Page 6: ...uring from then on will be displayed relatively to the set tare values When the tare function is activated the arrow Tara appears in the display To deactivate tare function press Tara for 2 seconds Pl...

Page 7: ...on 7 5 1 Unit Choice Of The Display Unit Choose the desired display unit the referring unit is displayed by means of a functional arrow in the display The selection is permanently stored in the sensor...

Page 8: ...r deviations Input is in the display unit 7 5 6 SCAL Adjusting Sensor Scale The scale of the measuring can be influenced by this setting factor is in displayed value measured value 1 Scal 100 Standard...

Page 9: ...l dia of 4 mm to pressure port B Port A will not be used Pressure sensors HND PS01 PS02 and P03 allow for measurements of under pressure up to the entire over pressure measuring range by re plugging t...

Page 10: ...thout battery wrong voltage or polarity Check power supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufactu...

Page 11: ...nge unit 7 Read max value 204 Read display range decimal point 12 Read ID number 208 Read of channels 176 Read min measuring range 214 Read Scale adjustment 177 Read max measuring range 216 Read Offse...

Page 12: ...ower off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7mm high for measuring values and for min max memories hold functio...

Page 13: ...0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 350 0 mbar rel 0 1 mbar max 1 4 bar 0 70 C G male stainless steel HND PS09 0 1000 mbar abs 1 mbar max 4 bar abs HND PS10 0 3500 mbar abs max 14 bar abs...

Page 14: ...face converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with lo...

Page 15: ...ectrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic product...

Reviews: