DOG-4
DOG-4 K08/0818
page 27
Electronic Options DOG-…M0/N0/O0/P0
(Transducer without/with ATEX certification and Flow computer)
Display:
Alphanumeric LCD, UV-resistant with
Displayed
functions:
Compensated Flow rate
(7 digits, 17 mm high)
Compensated total:
(7 digits, 17 mm high)
resettable
Accumulated
total
(11 digits, 8 mm high)
not
resettable
Actual line temperature
(6 digits)
Actual line pressure
(6 digits)
Units:
Flow:
m
3
, cf, scf, Nm
3
Time units: /sec, /min, /hr, /day
Total:
m
3
Accumulated total:
m
3
Temperature:
°C, °F, or K
Pressure:
mbar, bar PSI
Decimals:
Flow:
0, 1, 2 or 3
Total:
0, 1, 2 or 3
Accumulated total:
according to selection for total
Temperature/Pressure:
1
Backlightning: yes
Signal input:
Flow:
DOG-4 sensor
Temperature:
PT100, 2- or 3-wire
Pressure:
0(4)…20 mA (passive),
14-Bit resolution, 2- or 3-wire
Power supply:
M/N:
230 V
AC
± 10 %, 50…60 Hz
O/P:
110 V
AC
± 10 %, 50…60 Hz
Electrical Connection:
5 x M16 x 1.5 cable gland
Housing material:
ABS with PC cover
Weight:
approx. 1800 g
Analogue output:
4…20 mA (active),
10-Bit resolution, 3-wire
Pulse output:
PNP, 24 VDC active max. 50 mA,
scaled according to linearised accumulated total
(e.g. pulse every 12 liters)
Pulse duration: user defined
0.001 s…10 s
max. frequency 500 Hz
Protection: IP65
Mounting: Wall
mounting
Data protection:
EEPROM backup, backup of running totals every
minute, Data retention at least 10 years
Communication:
Modbus RTU RS485 2-wire
(optional, other Modbus versions on request)
Summary of Contents for DOG-4
Page 35: ...DOG 4 DOG 4 K08 0818 page 35 19 ATEX Certificate ...
Page 36: ...DOG 4 page 36 DOG 4 K08 0818 ...
Page 37: ...DOG 4 DOG 4 K08 0818 page 37 ...
Page 38: ...DOG 4 page 38 DOG 4 K08 0818 ...
Page 39: ...DOG 4 DOG 4 K08 0818 page 39 ...
Page 40: ...DOG 4 page 40 DOG 4 K08 0818 ...
Page 41: ...DOG 4 DOG 4 K08 0818 page 41 20 IECEx Certificate ...
Page 42: ...DOG 4 page 42 DOG 4 K08 0818 ...
Page 43: ...DOG 4 DOG 4 K08 0818 page 43 ...
Page 44: ...DOG 4 page 44 DOG 4 K08 0818 ...