Quickstart
Keypad
cal
Start, end calibration
conf
Start, end configuration
• Select digit position
(selected position blinks)
• Menu
navigation
• Edit
digit
• Menu
navigation
enter
• Calibration:
Continue in program sequence
• Configuration: Confirm entries,
next configuration step
• Measuring mode: Display output current
cal
enter
Cal Info: Display of zero and slope
conf
enter
Error Info: Display of last error message
+
Start GainCheck device self-test
Passcodes
Key + passcode
Menu item
cal + 0000
CAL info
(display of zero,slope)
cal + 1001
Zero calibration
cal + 1100
Calibration
(water / air)
cal + 1105
Product calibration
(zero)
cal + 1015
Temp probe adjustment
Confi guration
Key + passcode
Menu item
conf + 0000
Error info
(display of last error, erase)
conf + 1200
Confi guration
conf + 2222
Sensor monitor
(sensor current)
conf + 5555
Current source 1
(specify output current)
conf + 5556
Current source 2
(specify output current)
conf +
+ 4321
Factory setting