144
Menu Display
Cal timer, Tolerance band recorder
Calibration Timer
After expiration of a presettable inter-
val (Parameter setting, Module PHU,
Cal preset values), the calibration timer
generates a warning message as a
reminder that calibration is required.
The remaining time can be indicated
in the measuring mode by pressing a
softkey (secondary display: “CTime”).
Diagnostics of PHU 3400(X)-110
Calibration timer, Tolerance band recorder, Cal record, Sensor network diagram,
Statistics
Tolerance Adjustment
Records the tolerance ranges for zero
and slope over the time. If the values
determined by a calibration exceed
the tolerance limits, an adjustment
can be executed automatically.
Display can be graphical or as a listing.
The tolerance band for zero and slope
is configured during parameter setting
(Module PHU 3400-110,
Cal preset values).
Adaptive cal timer
22.3 °C
7.00 pH
T-band recorder
Return
Zero
22.3 °C
7.00 pH
Slope
Return
Next calibration in approx. 102 h
(02.04.10 14:08)
Default
Adaptive
Current
El stress
168 h
160 h
58 h
5 %
61.7
50.7
8.00
6.00
pH
°C
CTime 160h
Favorites menu
Adaptive Calibration Timer
The time until the next due
calibration is automatically reduced
depending on the electrode stress
(temperature, pH value).
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...