143
Menu Display
Module diagnostics /
Sensor monitor
Open diagnostics
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
confirm with enter.
Then select "Module PHU".
The Diagnostics menu gives an over-
view of all diagnostics functions avail-
able. Messages set as “Favorite” can
be called directly from the measuring
mode using a softkey.
To configure:
Parameter setting / System control /
Function control matrix.
Module Diagnostics
Internal function test.
Sensor Monitor
Shows the values currently measured
by the sensor. Important function for
diagnostics and validation!
Menu selection
Return to meas
22.3 °C
Select:
[enter]
Lingua
Module PHU 3400-110
22.3 °C
Return
Cal record
Sensor monitor
7.00 pH
7.00 pH
Sensor network diagram
Statistics
Set favorite
Module diagnostics
Sensor monitor
Return
pH input
22.3 °C
ORP input
7.00 pH
Temperature
Impedance glass (25°C)
Impedance ref (25°C)
-56 mV
200 mV
25.6 °C
100 MΩ
2.5 kΩ
Diagnostics of PHU 3400(X)-110
Module diagnostics / Sensor monitor
RTD
109.6 Ω
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...