105
Display text
Time [s]
01: Rinse water ON
0010 sec
02: Rinse water OFF
0002 sec
03: Probe in MEASURE
0005 sec
04: Meas duration
0030 sec
05: Probe in SERVICE
06: Cleaning agent
0020 sec
07: Rinse water ON
0060 sec
08: Rinse water OFF
0002 sec
09: Program end
Parameter Setting: Program Flows
Measuring (Short-Time)
Program step
Probe: Measure
Probe: Service
Rinse water
Cleaning agent
(Pump III)
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
07
08
Stroke 1
Stroke 2
Exposure time
Probe in MEASURE
(short-time)
t
[s]
Message (display)
Measurement (short-time) can also be started via a DCS input signal at input BIN1
of the Uniclean 900 probe controller.
Measuring program
(short-time)
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...