27
Cond
Measurement
cal
CAL
CELL CONST.
The conductivity will be shown in the display
and can be compared with a reference solution
(temperature‑corrected).
Value blinks
Use
to set the cell constant value.
cal
Calibration is performed. Display: CELL CONSTANT.
The meter then automatically returns to measuring mode.
CELL CONST. Calibration
(Calibration by entry of cell constant)
The calibration method is selected in the configuration menu.
Measurement
cal
CAL
COND
Immerse sensor in solution.
Value blinks
Use
to set the temperature‑corrected con‑
ductivity value.
Note
: Here, the meter does not
perform a temperature compensation!
cal
Calibration is performed. Display: CELL CONSTANT.
The meter then automatically returns to measuring mode.
COND Calibration
(Calibration by entry of conductivity)
The calibration method is selected in the configuration menu.
Conductivity Calibration
Summary of Contents for Portavo 904X COND
Page 2: ......
Page 67: ...67 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 68: ...68 pH Buffer Tables 02 Knick CaliMat Nominal values in bold C pH ...
Page 69: ...69 pH Buffer Tables 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated ...
Page 70: ...70 pH Buffer Tables 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added ...
Page 72: ...72 pH Buffer Tables 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH ...
Page 73: ...73 pH Buffer Tables 07 WTW Technical Buffers Nominal values in bold C pH ...
Page 74: ...74 pH Buffer Tables 08 Hamilton Nominal values in bold C pH ...
Page 75: ...75 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 76: ...76 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...