Specifications
Sensor standardization
*)
pH calibration
Operating modes
*
)
CALIMATIC
Calibration with automatic buffer recognition
MANUAL
Manual calibration with entry of individual
buffer values
DATA INPUT
Data entry of zero and slope
Calimatic buffer sets
*
)
-01- Mettler-Toledo
2.00/4.01/7.00/9.21
-02- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- NIST technical
1.68/4.00/7.00/10.01/12.46
-05- NIST standard
1.679/4.006/6.865/9.180
-06- HACH
4.01/7.00/10.01/12.00
-07- WTW techn. buffers 2.00/4.01/7.00/10.00
-08- Hamilton
2.00/4.01/7.00/10.01/12.00
-09- Reagecon
2.00/4.00/7.00/9.00/12.00
-10- DIN 19267
1.09/4.65/6.79/9.23/12.75
-U1- (User)
loadable via Paraly SW 112
Permissible calibration range
Zero point
6 ... 8 pH
With ISFET:
Operating point
(asymmetry)
-750 … +750 mV
Slope
approx. 74 … 104 %
(possibly restricting notes from Sensoface)
Calibration timer
*)
Interval 1 … 99 days, can be switched off
Sensoface
Provides information on the sensor condition
Evaluation of
zero/slope, response, calibration interval
*) User-defined