pH Configuration
Prior to measurement, a configuration should be performed to match
the connected sensor and the desired measurement performance.
Furthermore, you can select the suitable calibration method.
The following table gives you an overview. Factory settings are
shown in bold print.
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing set.
Configuring
Measurement
“Setup” display
Select using arrow keys, confirm by pressing set
.
Display 1
pH x.xx | pH x.xxx | mV
Display 2
OFF | date + time | date | time
CAL Timer
OFF | 1 ... 99 days
CAL
CALIMATIC | Manual | DATA INPUT |
(ISFET-Zero)
|
CAL SOP (Opt.) | ORP OFFSET (for pH/ORP combo sensor) |
TEMP. OFFSET (Opt.) | FREE CAL
CAL POINTS
1 | 2 | 3 | 1-2-3 (for CALIMATIC, Manual, FREE CAL)
BUFFER SET
(CALIMATIC,
FREE CAL)
-01- Mettler-Toledo
2.00|4.01|7.00|9.21
-02- Knick CaliMat
2.00|4.00|7.00|9.00|12.00
-03- Ciba (94)
2.06|4.00|7.00|10.00
-04- NIST technical
1.68|4.00|7.00|10.01|12.46
-05- NIST standard
1.679|4.006|6.865|9.180
-06- HACH
4.01|7.00|10.01|12.00
-07- WTW techn. buffers
2.00|4.01|7.00|10.00
-08- Hamilton
2.00|4.01|7.00|10.01|12.00
-09- Reagecon
2.00|4.00|7.00|9.00|12.00
-10- DIN 19267
1.09|4.65|6.79|9.23|12.75
-U1- loadable via Paraly SW 112 (User)
Auto OFF
OFF/ 0.1h / 1h / 6h / 12h
Temp Unit
°C / °F
Time Format
24h / 12h
Date Format
dd.mm.yy | mm.dd.yy
TAN TEMP CAL
(TAN input required, option; see page 37)
TAN SOP
(TAN input required, option; see page 37)
Setup Code
OFF (0000) | 0001 ... 9999 (with 001-SOP option only)
Cal. Code
OFF (0000) | 0001 ... 9999 (with 001-SOP option only)
Default
NO | YES (reset to factory settings)
Note: All data logger entries will be deleted.