10
Protection against
electric shock
3)
Protective separation to EN 61140 by reinforced insulation
according to EN 61010-1 up to 600 V AC/DC across input
and output for overvoltage category II and pollution
degree 2, up to 300 V AC/DC across output and ower
supply for overvoltage category II and pollution degree 2.
For applications with high working voltages, take measures
to prevent accidental contact and make sure that there
is sufficient distance to adjacent devices or sufficient
insulation between them.
EMC
4)
2004/108/EC
EN 61326
EN 61326/A1
Surge withstand
5 kV, 1.2/50 μs, to IEC 255-4
Ambient temperature
Operation
Transport and storage
-10 to +70 °C
-40 to +85 °C
Ambient conditions
Stationary application, weather-protected
Relative air humidity 5 ... 95 %, no condensation
Barometric pressure: 70 ... 106 kPa, altitude up to 2000 m
Water or wind-driven precipitation (rain, snow, hail)
excluded
Power supply
22 to 230 V AC/DC ± 10 %, 0.9 W, AC 48 to 62 Hz; 2.5 VA,
(overvoltage category II)
Construction
Modular housing with pluggable screw clamp terminals,
see dimension drawing
Protection
IP 20
Weight
approx. 150 g