68
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_RTD
calibration method.
Press enter to proceed.
Wrong settings change
the measurement
properties!
Measure the temperature of
the process medium using
an external thermometer.
Display (3 sec)
Now the device is in
HOLD mode.
Enter the measured tem-
perature value.
Maximum difference: 10 K.
Press enter to proceed.
Display of actual
temperature (uncom-
pensated) in the lower
display.
The corrected temperature
value is displayed.
Sensoface is active.
To exit calibration:
Select MEAS, then enter
To repeat calibration:
Select REPEAT, then enter
After end of calibration,
the outputs remain in
HOLD mode for a short
time.
After calibration is ended,
the device will switch to
measuring mode.
Temp Probe Adjustment