Diaphragm liquid pump NF 30, NFB 30
Installation and connection
KNF Flodos AG BA_NF30_EN_05_069206.docx
Translated from the Original Operating and Installation Instructions
15
➔
For pumps with fan: mount the pump so that the fan of the
motor can suck in sufficient cooling air.
➔
Make sure that the installation location is dry and the pump is
protected against rain, splashes, hose and drip water.
➔
Protect the pump against dust.
➔
Protect the pump against vibrations and jolts.
➔
Generally speaking, the pump can be mounted in any
orientation. The venting and accuracy of the pump is optimal if
installed as shown in the illustration (Fig. 12).
➔
KNF recommends mechanically decoupling the pump from the
pipework system, e.g. by using flexible hoses or pipes. This
prevents any oscillations of the pump being transferred to the
system.
Cooling air supply
Installation location
Mounting orientation
Fig. 13: Optimal mounting
orientation / horizontal
Decoupling