Diaphragm Liquid Pump NF 1.30
For your notes
KNF Flodos AG BA_NF1.30_EN_04_067674.docx
Translation of original Operating and Installation Instructions
9
6. Installation and connection
Install the pumps only under the operating parameters and condi-
tions described in Chapter 4. Technical data.
Observe all safety precautions (see Chapter 3).
6.1. Installation
Before installation, store the pump at the installation location to
bring it up to ambient temperature.
Mounting dimensions (see Fig. 3)
Fig. 3: Mounting dimensions NF 1.30 DCG
Make sure that the installation location is dry and the pump is
protected against water in the form of rain, spray, splashes and
drips.
Protect the pump against dust.
Protect the pump against vibration and impact.
Generally speaking, the pump can be installed in any position.
For maximum precision and rapid venting, install pump as
shown in the illustration (Fig. 4).
KNF recommends mechanically decoupling the pump from the
piping system, e.g. by using flexible hoses or pipes. This pre-
vents any oscillations of the pump being transferred to the sys-
tem.
Mounting dimensions
Installation location
Installation position
Fig. 4: Optimum
installation position
Decoupling