Maintenance and Care
39
AZURA V 2.1S User Manual, V6860, Version 2.2
Note:
If it is not possible to rectify the fault based on this list, then contact the
Technical Support.
Maintenance and Care
Cleaning and Caring for the Device
Device defect
Intruding liquids can cause damage to the device.
Place solvent bottles next to the device or in a solvent tray.
Moisten the cleaning cloth only slightly.
All smooth surfaces of the device can be cleaned with a mild, commercially
available cleaning solution, or with isopropanol.
Valve Assembly
The valve contains a rotor in its housing. With disk springs, the rotor creates a
certain pressure stability.
E4
Valve position 1 is not recog-
nized.
Readjust sensor board.
E5
Switch position of DIP 1 and 2
are not correct.
Correct DIP switch 1 and 2.
Binary code is not correct.
Check if valve setting (DIP
switches) is correct,, then send
the correct position parame-
ter (binary code).
E6
Memory error
Switch the device off and on
again.
Display
Cause Of the Fault
Solution