Installation
23
AZURA V 2.1S User Manual, V6860, Version 2.2
Inserting the Needle Seal
You fill the sample loop with a syringe or a feed pump. To do so, insert a needle
seal into the syringe connection and fixate it with a biconical seal (see “Seals”
on page 10). The needle seal for the syringe connection is available as PTFE or
Teflon tube.
Component defect
Rotor-seal damage caused from non-fitting seal.
Only reuse seals in one and the same location.
Preferably use a new seal.
Result
The 2-position valve is prepared for the injection of sample solution into the
sample loop via a glass syringe with luer lock.
Connecting the Device to the Computer
This section describes how to set up an HPLC system in a local area network
(LAN) and how a network administrator can integrate this LAN into your com-
pany network. The description applies to the operating system Windows® and
all conventional routers.
Note:
To set up a LAN, we recommend to use a router. That means the follow-
ing steps are required:
Process
1. On the computer, go to the control panel and check the LAN properties.
Process
Figure
1. Insert the needle seal
1
into the
Dynaseal fitting
2
.
2. Put the biconical seal
3
onto the
needle seal.
3. Guide the needle into the needle
seal , so that both ends fit equally.
4. Insert the needle seal and needle
into the port 1 and screw in manu-
ally. The Dynasel fitting sits ade-
quately tight when you feel a
resistance at pulling the needle.
Fig. 13 scheme of syringe port
1
2
3