30
Operation
AZURA® Detector UVD 2.1S Instructions, V6820
Adjusting V/AU ratio
By assigning different absorption values to a voltage value of 1 V, you can
stretch or compress the chromatogram. You can choose the absorption
value out of 4 values:
0.01 AU
0.1 AU
1 AU
10 AU
Status display is active.
Process
Figure
1.
Keep left arrow key pressed. Press right ar-
row key until the correct display appears.
2.
Let go of left arrow key.
3.
Using both arrow keys, set a value.
Fig. 19
Display
Analog-out
When setting is finished, the status display appears.
Display operating hours
Status display is active.
Process
Figure
1.
Keep left arrow key pressed. Press right ar-
row key until the correct display appears.
2.
Let go of left arrow key.
Fig. 20
Display lamp
operating hours
The number of operating hours is displayed and afterwards the status
display appears.
If the deuterium lamp has been operating for approximately 2000 hours
consider lamp replacement.
Using the analog port
To use the analog port for setting the wavelength of the detector, you
have to set a zero point and enter a scaling value.
Setting the zero point
The zero point is predefined by the applied voltage.
Device has been connected to GROUND.
ANALOG port has been selected.
Prerequisite
Procedure
A-OUT 1V
0.01AU
Result
Prerequisite
Procedure
1234.1h
GLP-lamp
Result
Next steps
Prerequisite