Installation and initial startup
AZURA® Detector UVD 2.1S Instructions, V6820
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
2
3
1
Fig. 8:
Operating tool
Process
1.
Insert the operating tool
3
in an upper small opening at the front of
the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted opera-
ting tool.
3.
Pull out the operating tool.
Check whether the cables are tightly attached.
Press the pin header onto the connector.
Finish the installation.
Bring the device into operation.
4.10.3 Analog control
Using the analog port, you can control the wavelength by changing the
applied voltage. A Mobile Control is required in order to select the option
ANALOG in the SETUP menu.
To use the analog port for controlling the detector, you have to set a zero
point and enter a scaling value.
Zero point at 0 V = 000 nm
Scaling: 100 nm per Volt
If 5 V voltage is applied, the wavelength is 500 nm.
Next steps
Example