Technical Data
17
DG 2.1S user manual V6880, version
3.3
Device defect
Intruding liquids can cause damage to the device.
Place solvent bottles next to the device or in a solvent tray.
Moisten the cleaning cloth only slightly.
What is to be done if the degasser leaks?
If the degasser leaks, solvent can reach the motor and from there reach the
degasser exhaust and flow out. In this case, the automatic pump control will
switch the device off.
Inform the Technical Support of KNAUER.
Tightening Screw Fittings
1. If the screw fitting of the solvent tube is untight, carefully tighten the screw
fitting.
2. If it still leaks, loosen and check the screw fitting.
3. If no damage or deformation is visible at the seal ring and at the screw fit-
ting, repeat step 1.
4. If it still leaks, replace the screw fitting and seal ring.
Preventative Measures
To ensure that the degasser can constantly run at maximum capacity, pay
attention to the following:
Only use ultrapure solvent as mobile phase. If water is used as solvent, it also
has to fulfill this standard or be filtered and deionized.
Filter all solvents to prevent clogging.
To dry the degasser, only use ultrapure gas.
Observe the information on switching off the degasser.
Technical Data
Channels
2
Maximum flow rate
10 ml/min
Degassing method
gas permeation through a fluoropolymer membrane
Dead volume
approx. 285 μl per channel
Pressure decline
1.37 mm (Hg/ml/min)
Max. pressure stability
7 bar
Wetted materials
PEEK, glass-filled PTFE, Teflon AF
®
Display
1 LED
Dimensions
121 x 138 x 190 mm