background image

Technical Data

17

DG 2.1S user manual V6880, version 

3.3

Device defect

Intruding liquids can cause damage to the device.

Place solvent bottles next to the device or in a solvent tray.

Moisten the cleaning cloth only slightly.

What is to be done if the degasser leaks?

If the degasser leaks, solvent can reach the motor and from there reach the 
degasser exhaust and flow out. In this case, the automatic pump control will 
switch the device off.

Inform the Technical Support of KNAUER.

Tightening Screw Fittings

1. If the screw fitting of the solvent tube is untight, carefully tighten the screw

fitting.

2. If it still leaks, loosen and check the screw fitting.
3. If no damage or deformation is visible at the seal ring and at the screw fit-

ting, repeat step 1.

4. If it still leaks, replace the screw fitting and seal ring.

Preventative Measures

To ensure that the degasser can constantly run at maximum capacity, pay 
attention to the following:

Only use ultrapure solvent as mobile phase. If water is used as solvent, it also
has to fulfill this standard or be filtered and deionized.

Filter all solvents to prevent clogging.

To dry the degasser, only use ultrapure gas.

Observe the information on switching off the degasser.

Technical Data

Channels

2

Maximum flow rate

10 ml/min

Degassing method

gas permeation through a fluoropolymer membrane

Dead volume 

approx. 285 μl per channel

Pressure decline

1.37 mm (Hg/ml/min)

Max. pressure stability

7 bar

Wetted materials

PEEK, glass-filled PTFE, Teflon AF

®

Display

1 LED

Dimensions

121 x 138 x 190 mm

Summary of Contents for Azura Degasser DG 2.1S

Page 1: ...Degasser DG 2 1S Instructions Document No V6880 HPLC...

Page 2: ...Note The contents of this user manual apply for the BlueShadow product Please submit a request on any article numbers for BlueShadow...

Page 3: ...king and Setup 11 Contacting the Technical Support 11 Location Requirements 11 Unpacking 12 Initial Startup 12 Screw Fitting of the Solvent Tube 12 Solvent Tube Connection 12 Connecting Degasser and P...

Page 4: ...Contents DG 2 1S user manual V6880 version 3 3 Technical Data 17 Accessories and Spare Parts 18 Legal Information 18 Transport Damage 18 Warranty Conditions 18 Declaration of Conformity 19 Disposal 1...

Page 5: ...outlet side one bicolor LED and the exhaust port Air and possibly solvent vapors will be pumped out through the exhaust port If necessary it can be combined with a fume hood or similar extraction dev...

Page 6: ...nd routine lab oratories Features Reliable and convenient degassing of up to two eluents at the same time Teflon AF allows high gas diffusion rates Constant performance vacuum pump for high baseline s...

Page 7: ...by the manufacturer If there is any doubt contact the technical support of the manufacturer Not suitable eluents Chlorobenzene Chloroform Concentrated acids Concentrated bases Eluents containing part...

Page 8: ...installation of a device or a training by the company KNAUER or an authorized company If you do not belong to this or a comparable professional group you may not perform the work described in this use...

Page 9: ...d to connect the device and the power supply system self ignition point Only use eluents that have a self ignition point higher than 150 C under normal ambient conditions power strip If several device...

Page 10: ...technically correct manner All contaminated devices must be properly decontaminated by a specialist com pany or the operating company before they can be recommissioned repaired sold or disposed of Al...

Page 11: ...evice overheats at exposure to sunlight and insufficient air circulation Device failures are very likely Set up the device in such a way that it is protected against exposure to direct sunlight Keep a...

Page 12: ...ing contact the technical support 5 Grip the device at its side panels and lift it out of the packaging 6 Remove the foam inserts from the device 7 Check for damages caused during transportation In ca...

Page 13: ...egassing chambers with screw fittings and seal rings These tubes are then connected to the respective pump Switching On the Degasser Requirements All lines have been connected correctly Unused degasse...

Page 14: ...If a leak is detected the microprocessor of the degasser automatically increases the pump speed to maintain the vacuum If the vacuum still falls below the target value the yellow LED flashes If it is...

Page 15: ...by means of laboratory gas 5 Seal all inlets and outlets of the degasser with blind fittings and subsequently store the device at a safe and dry location 6 Prior to the reuse of the degasser purge th...

Page 16: ...t hardly hear the pump At low rotational speeds the pump can hardly be heard although the vac uum is good and the degassing process is tak ing place Compare the baseline of non degassed ethanol at 251...

Page 17: ...mage or deformation is visible at the seal ring and at the screw fit ting repeat step 1 4 If it still leaks replace the screw fitting and seal ring Preventative Measures To ensure that the degasser ca...

Page 18: ...ranty period any components with material or design related defects will be replaced or repaired by the manufacturer free of charge Please connect to our website for further information on terms and...

Page 19: ...ment 160214 WEEE Registration KNAUER as a company is registered by the WEEE number DE 34642789 in the German Elektroaltger teregister EAR The number belongs to category 8 and 9 which among others comp...

Page 20: ...lation location see location 11 L leak 9 Leak test 14 Leakage 17 Location 11 Long term switch off 15 M Modifiers 7 O Order numbers 18 P power cable 9 power strip 9 power supply 9 professional group 8...

Page 21: ...tions online https www knauer net en Support User manuals KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Phone 49 30 809727 0 Fax 49 30 8015010 E Mail info knauer net Internet www kn...

Reviews: