ASM 2.1L user manual V6815
17
Initial startup
Leak management
The leak management consists of the leak sensor and the drainage system
(funnels, hoses, nozzles). The drainage system ensures that escaping liq-
uids flow into a waste bottle. When leaks are registered by the leak sen-
sor, the LED flashes red. Both the device and the data acquisition via
chromatography software are stopped.
Prerequisite
The front cover has been removed.
Next steps
Attach the front cover.
Process
Process
Figure
1. Carefully push the funnel
1
into the center opening of the
capillary guide
2
.
Fig. 7
Funnel and capillary guide
2. Push the long ending of the first
nozzle
4
into the hose
3
.
Fig. 8
Hose and nozzle
3. Connect the nozzle and the fun-
nel.
4. Push the other end of the hose
onto the nozzle
5
of the leak
tray.
Fig. 9
Hose connected to device
5. For the bottom device, push
the short end of the nozzle
6
into the opening in the collec-
tion point of the leak tray.
6. Connect the hose to the nozzle
and lead the second ending to
the waste bottle.
7. Place the waste bottle below
the bottom device.
Fig. 10
Leak tray with nozzle
1 2
3
4
5
6