DAD 6.1L/DAD 2.1L/MWD 2.1L user manual V6700, version 3.2
46
Technical Data
Next Steps
Bring the device into operation.
Removing the Leak
Prerequisites
If the leak tray is filled with liquid, the leak has to be removed.
Auxiliary Material
Cloth for drying the leak sensor
Next Steps
Bring the device into operation.
Technical Data
DAD 6.1L
4. Using the screwdriver, tighte the
two screws
4
or
7
on the lamp
socket.
5. Connect the plug
5
or
6
and
tighten the lock ring.
Fig. 37
installing the deuterium lamp
6. Attach the lamp cover and fix it
with the screws
8
and
9
.
Fig. 38
lamp cover
Process
Process
Figure
7
6
4
5
8
9
Process
Process
1. Remove the leak.
2. Dry the leak tray.
3. Acknowledge the error message via Mobile Control, Control Unit, the
software or the standby switch (without software control).
Detection
detector type
diode array detector
number of diodes
1024
resolution of diodes
0.8 nm/diode
detection channels
8 (digital), 4 (analog)
light source
high intensity deuterium (D
2
) lamp and halogen
lamp with integrated GLP chip
wavelength range
190–1000 nm
spectral bandwidth
< 3.5 nm at H
α
line (FWHM)
Note:
digital bandwidth 1–32 nm
wavelength accu-
racy
± 1 nm