30
Revision 1
4.5
Table Charging Unit with 4 Chambers
The table charging unit with 4 chambers is intended for keeping fully charged replacement
batteries ready for use.
The batteries are fully charged after approx. 10 h.
•
Position the console in a safe position on a table top that is protected from dangers of
spillage and fire near a wall box.
•
Insert the charging cable.
—
The red/orange LEDs light up.
•
Insert the batteries into the chambers.
Ensure that the batteries are properly aligned and fully inserted into the chamber.
Indicator
Meaning
Green flashing light
Battery is being charged
Green continuous light
Battery is fully charged
Red/orange continuous light
No battery inserted or no contact.
Fully re-insert the battery.
Red flashing light
Battery defective.
Try another battery.
5
Cleaning
•
Before cleaning, disconnect the charging unit from the headlight and power supply.
The device casing can be cleaned using a soft cloth that has been moistened with alcohol or a
non-aggressive disinfectant solution. The black padding strips are cleaned using the same
method. The padding strips can be removed for cleaning by opening the pushbuttons.
However, we recommend leaving them on the headlight for the cleaning.
•
Avoid excessive moisture, as this can lead to damage on the electrics, safety risks and
voidance of the warranty.
•
Do not immerse in or spray with disinfection agents.
New padding strips are available as accessories.
Summary of Contents for MedLED
Page 2: ...2 Revision 1 Deutsch 6 English 21 Español 36 Français 51 Italiano 66 ...
Page 81: ...Istruzioni per l uso MedLED Lampada frontale Revision 1 81 ...
Page 82: ...Istruzioni per l uso MedLED Lampada frontale 82 Revision 1 ...
Page 83: ...Istruzioni per l uso MedLED Lampada frontale Revision 1 83 ...