Page 1: ...ront Cover English independant Fix Bridge independant Fix Ceiling independant Fix Wall Instructions for Use 0051 90 577 52 10 Revision 01 Date of Release 2020 11 www klsmartin com...

Page 2: ...ould result in death or serious injury DANGER Indicates a situation which if not avoided will result in death or serious injury Instructions for Use Serial number Date of manufacture Manufacturer Dist...

Page 3: ...pplicable SOCKET 1 SOCKET 2 Socket identification as per electrical diagram where applicable xxxVA lighting Power consumption of each light Correct Configuration Key required Inlay shape diameter Clic...

Page 4: ...ion 9 5 Screw Torques 10 6 Functioning 11 7 General and Technical Features 16 7 1 General Features 16 7 2 Technical Specification 16 8 Cleaning and Maintenance 18 8 1 Required Maintenance 18 8 2 Recom...

Page 5: ...dle the device or product or use it for clinical applications please do not hesitate to contact the Product Management Phone 49 7461 706 345 Fax 49 7461 706 190 In case of technical questions and ques...

Page 6: ...DA installation must be done following indication listed on manufacturer installation manual Electrical connection must be done following the manufacturer electrical scheme Gas outlet distribuiton mus...

Page 7: ...d on the plate Please don t connect adapters between the device s electrical sockets and plugs Risk of injury due to system parts falling down Please don t overload UDA s max load listed on the relate...

Page 8: ...other equipment 3 Intended Use The pendant models Indepedant Fix are manufactured for medical use areas only in order to allow Electrical current flow Data nurse calling system phone PC network Medica...

Page 9: ...independant Fix Instructions for Use Revision 01 9 4 Configuration independant Fix Bridge independant Fix Wall independant Fix Ceiling...

Page 10: ...independant Fix Instructions for Use 10 Revision 01 5 Screw Torques Dimension Torques M5 7 Nm M6 12 Nm M8 30 Nm...

Page 11: ...independant Fix Instructions for Use Revision 01 11 6 Functioning...

Page 12: ...independant Fix Instructions for Use 12 Revision 01...

Page 13: ...independant Fix Instructions for Use Revision 01 13...

Page 14: ...independant Fix Instructions for Use 14 Revision 01...

Page 15: ...independant Fix Instructions for Use Revision 01 15...

Page 16: ...est For each unit the following tests are performed Ground impedance protection conform to standard 8 6 laid down by IEC EN 60601 1 Standard leakage current and dielectric strength measure conform to...

Page 17: ...component parts lists calibration instructions or any other useful information that technical assistance may requires in order to repair component s replaceable parts Manufacturer registered office an...

Page 18: ...EN ISO 7396 1 and EN ISO 7396 2 8 2 Recommended Maintenance We recommend to check the following list every 6 months Correct operation of the brake system LEDs movement and braking correct Possible mo...

Page 19: ...o not overlap more than 5 devices and do not place them vertically In case of irreparable damages or end of life equipment the device must be disposed following the existing rules If the user does not...

Page 20: ...es Independent Fix is easy to clean due to smooth surfaces Is Independent Fix easy to maintain No maintenance is needed except for gas implants please refer to this manual Does Independent Fl Fix ex i...

Page 21: ...ve humidity should be at least 30 Electrical fast transient 2 kV phase i Earth 1 kv phase i phase i 2 kV common mode 1 kv differential mode Voltage quality should be like a typical hospital environmen...

Page 22: ...tance in metres Radiowave radiated power 3 V m 80 MHz to 2 GHz 3 V m P d 35 0 80 MHz to 800 MHz P d 7 0 800 MHz to 1 GHz P d 33 2 1 GHz to 2 5 GHz where P is the maximum output power rating of the tra...

Page 23: ...or Use Revision 01 23 12 Maintenance Log Book Serial number MD Official distributor Required maintenance Frequency Registration Name Date Inspection of flexible gas hoses Each 6 months Replacing of fl...

Page 24: ...sia Tel 7 499 792 76 19 russia klsmartin com KLS Martin Taiwan Ltd Taipei Taiwan Tel 886 2 2325 3169 taiwan klsmartin com Gebr der Martin GmbH Co KG Dubai United Arab Emirates Tel 971 4 454 16 55 midd...