Table 203
Evidence of
immunity
Test level
IEC 60601
Level of
compliance
Electromagnetic environment
(distance of transmitter equipment)
Conducted radio
frequency
3 Vrms
3 Vrms
P
d
⋅
=
17
.
1
where
P
means the maximum output power rating in Watt of the
transmitter and
d
is the separation distance in meters
Radio wave
radiated power
3 V/m
(80 MHz to 2 GHz) 3 V/m
P
d
⋅
=
35
.
0
80 MHz to 800 MHz
P
d
⋅
=
7
.
0
800 MHz to 1 GHz
P
d
⋅
=
33
.
2
1 GHz to 2.5 GHz
where
P
is the maximum output power rating of the transmitter and
d
is the separation distance in meters
Table 204
Maximum output
power (W) transmitter
Separation to the frequency of the transmitter (m)
15 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 1 GHz
1 GHz to 2.5 GHz
0.01
0.12
0.04
0.07
0.23
0.1
0.37
0.11
0.22
0.74
1
1.17
0.35
0.70
2.33
10
3.69
1.11
2.21
7.38
100
11.67
3.50
7.00
23.33