Initial Operation 

 

2.

 

Initial Operation 

With its integrated NiMH storage battery the CPM 374 is plug-and-play. After switching on the 
software version is displayed briefly. 

Please use only the provided power supply. In battery operation mode the backlight is powered 
off to save power. Furthermore the device has automatic power shut-off after 10min of inactivi-
ty.  

3.

 

Mode Selection 

The CPM 374 can be driven as charge plate monitor, voltmeter or electrostatic field meter. By 
pressing button 

«B» 

the modes can be switched. 

 

4.

 

Charge Plate Mode 

4.1.

 

Decay Time and Offset Voltage 

The display shows the selected type of measurement and prompts to start, e.g.: 

 

The last mode is stored in the EEPROM and displayed after turning on. 

By selecting 

Decay Time = AUTO 

the decay time for the positive and the negative voltage  is de-

termined automatically . 

AUTO

 is recommended because the Start button has to be pressed only 

once to perform a measurement, including the determination of the offset voltage. 

The 

offset voltage 

is the plate’s potential after a certain time that defines the equilibrium of 

positive and negative ions. If the offset voltage is positive, the air ionizer has a excess of positive 
ions – in case of negative offset voltage the air ionizer excesses negative ions. 

Please consider the manufacturer’s instructions for a qualitative analysis. 

After pressing the button 

«A»,

 the plate electrode is charged by the integrated high voltage unit. 

The display shows the timer parameters and the applied voltage, e.g.: 

 

TIMER: 1000-100V 

U > 1200V 

DECAY TIME: AUTO 

PRESS START ! 

Summary of Contents for CPM 374

Page 1: ...MANUAL Charge Plate Monitor CPM 374 Incl Electrostatic Field Meter System Prof Dr Ing Hans Kleinw chter...

Page 2: ...Setup 8 4 3 2 Calibration 10 5 Voltmeter Mode 11 5 1 Voltage Metering 11 5 2 Setting up the Measurement Range 11 5 3 Settings 11 5 3 1 Change distance 11 5 3 2 Zero Adjust 12 6 Electrostatic Field Me...

Page 3: ...ply As the CPM 374 is menu driven and operated by 2 pushbuttons the device is very user friendly For a better orientation the selected parameters are displayed in the double spaced LCD display The dev...

Page 4: ...up 1 3 2 4 6 5 11 13 12 8 7 10 9 1 Plate electrode 8 Handhold 2 LCD Display 9 Mains connection 12V DC 3 Button A 10 Serial interface 4 Protective cap 11 1V analog output with banana jack 5 Modulator s...

Page 5: ...the EEPROM and displayed after turning on By selecting Decay Time AUTO the decay time for the positive and the negative voltage is de termined automatically AUTO is recommended because the Start butto...

Page 6: ...offset voltage is measured after a defined time period The measurement is finished after determining the offset voltage All data required e g to quali fy an air ionizer can be read off the display The...

Page 7: ...To save the data nevertheless the memory has to be erased Therefore the following prompt is displayed Skip the deleting process by pressing button B and continue to the next measurement The last resu...

Page 8: ...3s the display returns and you can select the new value by pressing button A 4 3 Settings By pressing A and B simultaneously the device prompts after 1s Press A to enter SETUP or B followed by A to e...

Page 9: ...ards the time period for the offset voltage measurement is inquired Confirm the displayed value by pressing A or press B to select a new one Values are switched in the following order The selected par...

Page 10: ...ards Press button B to discard the changes The old parameters are set again and the changes that have been made are discarded 4 3 2 Calibration By pressing B followed by A after the SETUP question the...

Page 11: ...voltmeter mode the display shows By pressing A the following ranges can be set The display shows the current range in the example above R 2kV 5 3 Settings After pressing A and B simultaneously the dev...

Page 12: ...as to be grounded Ground the measuring head and confirm that by pressing button A Afterwards the zero adjust is performed and the device returns Important During the zero adjust external voltage must...

Page 13: ...choose between the follow ing options 6 3 Zero Adjust The provided protective cap has to be mounted on the modulator system to perform a zero ad just After pressing A and B simultaneously for about 1...

Page 14: ...nty In case of proper operation according to the manual we issue a guarantee of 24 months Mechan ical damage and battery destruction by wrong handling are excepted Guarantee expires on opening the dev...

Page 15: ...electrode is insulated it has to be protected from electrostatic induction Electrostat ic induction can be effected by a person passing by or by moving a charged object close to the instrument Extern...

Page 16: ...16 Manual CPM 374...

Reviews: