Kleinwächter EFM-122 Manual Download Page 5

 

5

4.6 

Change measurement distance 

Press and hold the „Distance“-key (with a screw driver through the hole on the right side) 
until „change distance“ will appear in the second line. The pre-selected distance in cm will be 
displayed in the first line. Press the „function/on“-key shortly to change the distance. With 
each short press of the „function/on“-key the proposed distance will appear as follows: 
 
2,5cm => 5cm => 10cm => 20cm =>cps - mode => 1cm  ... 
 
When the right distance is shown in the display, please wait (do not press the „function/on“- 
key) until the actual measurement value will appear in the second line. The new selected 
distance is now also shown in the first line. Please consider this selected distance for the 
following measurements. The distance will be measured from the electrode positioned 
6.4mm apart from the housing on the modulator side. It is more practical to add 6.4 mm to 
any selected distance starting from the front - plate of the unit. 

 

 

Please note: the distance will be stored  ! 
 
 
4.7 

Display 

The alphanumeric Liquid Crystal Display (LCD) consists of 2 lines for 12 digits each. The 
measurement distance in cm will be shown in the first line, while the voltage reading in volts 
is displayed in the second line. When the level of 999 volts is reached, the display will switch 
automatically into the kV-mode. The display of the voltage reading appears always in 3 
digits, such as: 578 V  3,85 kV  24.0 kV  and  274 kV. Select a greater distance if „overflow !“ 
appears on the screen ! 
 
 
4.8 

Error message 

Appears an error message in the display, the unit switch off after 2 seconds. If the same 
error message appears again after restart, please send the unit back for repair. 

 
 

4.9  

Battery voltage control 

The EFM 022 has a permanent battery-voltage control. If the battery voltage is less than 7,6 
V „Low Battery“ appears in the first line of the LCD and the 9V NiMH block-battery has to be 
charged or replaced ! 
 
Note: Please use rechargeable - NiMH 9V - batteries only!  
 
 
 
5. PC-Interface 

The EFM 122 has a serial RS 232 Interface. Connect the EFM with the included serial Cable 
to a PC. Put the 3,5” Disc into the PC and start “Install” 
Following the instructions on the PC. 
 
 

6. Grounding 

The unit has to be connected to ground for correct measurements of voltage levels and 
polarity. Use the grounding socket (4) of the unit for the connection to earth potential. The 
unit consists of a conductive housing, thus allow the use without grounding cable if the 
operator is on earth potential (e.g. wears conductive shoes, uses a wrist strap or touches 
grounded parts). 
 

Summary of Contents for EFM-122

Page 1: ...ATIC FIELD METER EFM 122 Small hand held Electro Fieldmeter with digital display and separate sensor to measure the electrostatic charge Display directly the electrostatic voltage potentials with pre...

Page 2: ...lication 4 4 3 Operating 4 4 4 Hold Function 4 4 5 Ranges 5 4 6 Changing the distance 5 4 7 Display 5 4 8 Error Messages 5 4 9 Battery Control 5 5 PC Interface 5 6 Grounding 5 7 Maintenance 6 7 1 Char...

Page 3: ...ge of the electrostatic field has to be located and the polarity and strength will be measured To do so as well as check any chosen steps against electrostatic charge or to control desire charge our E...

Page 4: ...electric charges electrostatic charges extremely high resistant voltage sources 4 3 Operating Press the function on key to switch on the unit Remove the protection cup from the modulator Press the fun...

Page 5: ...ding in volts is displayed in the second line When the level of 999 volts is reached the display will switch automatically into the kV mode The display of the voltage reading appears always in 3 digit...

Page 6: ...olerance 7 3 Guaranty Performance By correct operation and in accordance to the instruction manual we allow a 24 month warranty The warranty does not include The NiMH Battery damages by electrical dis...

Page 7: ...rface Alphanumeric LCD Display 2x12 function on Key I H Key to change the distance only to use with a screw driver Optional accessories Tripod for fixation of the measure sensor measure head MK11 with...

Page 8: ...8 Kleinw chter GmbH www kleinwaechtergmbh de...

Reviews: