5

200pF range and keep increasing until the over-range 

indication goes off and a reading is obtained. 

b) 

A shorted capacitor will read over-range at all range. A 

capacitor with low voltage leakage will read over-range, 

or a much higher value than normal. A open capacitor 

will read zero at all ranges (maybe a little reading at 

200pF range due to stray capacitance of the meter).   

c) 

When measuring the very low capacitance, it  should 

be performed using extremely short leads in order to 

avoid introducing any stray capacitance. 

d) 

When using the optioned test leads, remember that the 

leads maybe introduce a measurable capacitance to 

influence the accuracy of measurement. As a tip of 

measurement, the test lead capacitance may be 

measured by opening the leads and recording it down, 

then subtracting it to obtain a more accurate measuring 

result. 

e) 

Capacitors, especially electrolytic, it often have a wide 

tolerance. Do not be surprised if the measured value is 

greater than the value marked on the capacitor, unless it 

is a close close tolerance type. However, the value are 

seldom drastically lower than the rated value. 

f) 

If changing the range, measured value is changed, it 

shows that is a leakage-voltage capacitors. However, 

the influence of leakage-resistance will be decreased in 

lower range. 

Summary of Contents for KJ6013

Page 1: ...KJ6013 Digital Capacitance Meter Operation Manual ...

Page 2: ...tment manual adjusting range is 20pF Display Large size LCD display Max 1999 counts Range 9 ranges measuring range is from 0 1pF to 20mF Sampling time 0 5 seconds Calibrating adjustment a panel zero adjuster for zeroing adjustment Over range indication 1 display Low battery indicator display Operating humidity 80 RH Operating temperature 0 40 32 104 Storage temperature 10 50 14 122 Power supply si...

Page 3: ...ency 200pF 199 9pF 0 1pF 800Hz 2nF 1 999nF 1pF 800Hz 20nF 19 99nF 10pF 800Hz 200nF 199 9nF 100pF 800Hz 2µF 1 999µF 1000pF 800Hz 20µF 19 99µF 0 01µF 80Hz 200µF 199 9µF 0 1µF 8Hz 2mF 1 999mF 1µF 8Hz 20mF 19 99mF 10µF 8Hz Unit conversion formula 1Farad 10 3 Millifarads 10 6 Microfarads 10 9 Nanofarads 10 12 Picofarads Accuracyis pencentage of reading number of last digits at 23 5 80 RH 200pF 0 5 1 0 ...

Page 4: ...and terminal it will cause the battery dissipation rise This C meter is intended for measuring the capacitance value of capacitor It is not intended for determining the Q factor for above reactive components Misleading readings may be obtained if the measurement of capacitance of a resistor is attempted When measuring in circuit that circuit must be switched off and de energized before connecting ...

Page 5: ...e measured Confirm display show 0000 firstly low test range such as 200 pF 2 nF 20 nF need to adjust ADJ to be zero Connect the measured capacitor and press TEST DIS button to discharge observe polarity when connecting polarity capacitor Release TEST DIS button measuring Read the display the value is direct reading in the unit pF nF uF or mF indicated at the selected range switch If display show 1...

Page 6: ...tioned test leads remember that the leads maybe introduce a measurable capacitance to influence the accuracy of measurement As a tip of measurement the test lead capacitance may be measured by opening the leads and recording it down then subtracting it to obtain a more accurate measuring result e Capacitors especially electrolytic it often have a wide tolerance Do not be surprised if the measured ...

Page 7: ...ean the surface of the instrument Take out of the battery to prevent the battery electrolyte leakage if don t use for long time Battery replacement Replace the battery immediately when LCD indicates symbol steps as the following Take out of the screw and remove battery cover Remove 9V battery and replace with another new battery with same type or equivalent Reinstall the battery cover and fasten t...

Page 8: ...7 new one with same type or equivalent Reinstall the bottom case and fasten the screws End This operating manual subject to change without notice ...

Reviews: